 9.0  Troubleshooting  Guide

Problem

Possible Cause

Solution

Centrifuge will not start

No power supply

 



Check that power is being supplied to 
the outlet

 



Check that the power cord is plugged 
into both the wall

 



outlet and the back of the centrifuge

 



Check that power cord is not damaged

Blown fuse

Check fuse and replace if necessary

Lid lock will not release

Defective lid lock

Open manually and have unit serviced

No power from PC board

Call for service

Lid lock is jammed

Call for service

Centrifuge is not receiving 
power

See “Centrifuge will not start”

Centrifuge cannot be 
started, although power 
is on

Lid not closed correctly

Close lid correctly

No speed or time has been 
selected

Set speed and/or time

bAL: Error Message 
Indicates imbalance

Tubes not inserted 
symmetrically in rotor holes

Load tubes symmetrically (see Section 

3.4

 

on Loading the rotor)

Sample liquid in tubes not 
balanced

Make sure the same volume of liquid is in 
each tube

Defective or improperly 
adjusted balance sensor

Call service

Lid: (Error message) Lid 
not closed

Lid not closed completely

Close lid

Lid lock or sensor defective

Call service

Other Error Messages Er 
on Display

 



Press time or speed knob to clear error

 



Call service

Should you have a question about the operation of the Labnet Prism™ Microcentrifuge or if service is 
required, contact Customer Service. Do not send in a unit for service without first calling to obtain a 
repair authorization number. Should the unit require return for service, it should be properly packed to 
avoid damage. Any damage resulting from improper packaging shall be the responsibility of the user.

Содержание C2500

Страница 1: ...Prism Microcentrifuge Instruction Manual Catalog Numbers C2500 C2500 230V This manual is available in additional languages at www labnetlink com ...

Страница 2: ...ssories 3 6 2 Rotor Maintenance 4 6 3 Removing and Installing the Angle Rotor 4 6 4 Loading the Rotor 4 6 5 Overloading Rotors 4 7 0 Operation 5 7 1 Attaching Rotor Lid 5 7 2 Closing the Lid 5 7 3 Lid Release 5 7 4 Lid Lock 5 7 5 Speed Selection 5 7 6 Selection of Operating Time Momentary Operation Start Stop 6 8 0 Service and Maintenance 6 8 1 Centrifuge Service 6 8 2 Cleaning 6 8 3 Replacing Fus...

Страница 3: ...l equipment Never centrifuge flammable explosive or corrosive materials Never centrifuge hazardous materials outside of a hood or proper containment facility Always load the rotor symmetrically Each tube should be counter balanced by another tube of the same type and weight Always locate the centrifuge within easy access to an electrical outlet Always use only microcentrifuge tubes made from plast...

Страница 4: ...anual 1 each Warranty card 1 each Power cord 1 each 5 0 Installation The centrifuge should be installed on a rigid even surface such as a stable laboratory bench countertop etc To guarantee sufficient ventilation ensure that the centrifuge has at least 15 cm 6 inches of free space on all sides including the rear The centrifuge should not be located in areas subject to excessive heat such as in dir...

Страница 5: ...included Cat No AR110 Strip Spin Adapter Cat No C2400 SS Tubes Accepted 0 2 mL tubes or 8 x 0 2 mL strips Maximum Speed 15 000 rpm Centrifuge Radius Range 4 32 cm to 5 33 cm RCF G value Range 10 874 x g to 13 405 x g Adapter for 0 5 mL tubes pk 6 Cat No C 1205 Tube Measurement 8 x 30 mm Maximum Speed 15 000 rpm Centrifuging Radius 7 53 cm RCF G value 18 942 x g Adapter for 0 4 mL tubes pk 6 Cat No...

Страница 6: ...rotor wrench 6 4 Loading the Rotor Tubes to be loaded should be filled equally by eye The difference in the weight between the tubes should not exceed 0 1 gram Tubes should always be loaded so that there is equal spacing between all tubes One or two additional loaded tubes may need to be added to achieve this Refer to Figure 1 to see one typical balancing scheme 6 5 Overloading Rotors The maximum ...

Страница 7: ...WARNING Do not attempt to open the lid of any centrifuge until the rotor has come to a complete stop In the event of a power failure or malfunction it may be necessary to open the lid manually 1 Disconnect the power cord from the wall socket 2 Remove the plastic plug located on the left side of the unit 3 Pull the wire attached to the plug to open the lid lock manually 7 4 Lid Lock The centrifuge ...

Страница 8: ...housing rotor chamber rotor and rotor accessories clean All parts should be wiped down periodically with a soft cloth For more thorough cleaning use a neutral cleaning agent pH between 6 and 8 applied with a soft cloth Excessive amounts of liquid should be avoided Liquid should not come into contact with the motor After cleaning ensure that all parts are dried thoroughly by hand or in a warm air c...

Страница 9: ... and or time bAL Error Message Indicates imbalance Tubes not inserted symmetrically in rotor holes Load tubes symmetrically see Section 3 4 on Loading the rotor Sample liquid in tubes not balanced Make sure the same volume of liquid is in each tube Defective or improperly adjusted balance sensor Call service Lid Error message Lid not closed Lid not closed completely Close lid Lid lock or sensor de...

Страница 10: ...ge has an automatic g force conversion program so g values are automatically calculated and can be displayed on the control panel for this centrifuging radius If adapters or smaller tubes are used the centrifuging radius changes as does the g force The chart on the next page can be used to manually determine g values for any centrifuging radius ...

Страница 11: ...s or contact your local support office Corning s Customer Service team will help arrange local service where available or coordinate a return authorization number and shipping instructions Products received without proper authorization will be returned All items returned for service should be sent postage prepaid in the original packaging or other suitable carton padded to avoid damage Corning wil...

Страница 12: ...ur local Corning sales office www labnetlink com For a listing of trademarks visit www corning com clstrademarks All other trademarks are the property of their respective owners 2021 Corning Incorporated All rights reserved 4 21 CLSLN AN 1003DOC REV1 Corning Incorporated Life Sciences www corning com lifesciences NORTH AMERICA t 800 492 1110 t 978 442 2200 ASIA PACIFIC Australia New Zealand t 61 4...

Отзывы: