• When pipetting acids or corrosive solutions which emit

vapors, it is recommended to disassemble the shaft
and to rinse the piston and seal with distilled water after
finishing the pipetting operation.

8 - RECALIBRATION

Labnet

pipettes are calibrated by gravimetric method,

using 

Labnet

tips and distilled water, at the temperature

20±1°C, according to EN ISO 8655 standard.
If during pipette operation you find that the accuracy
error (the difference between the real aspirated volume
and the preset volume) exceeds the permissible value
given in the table in section 1, the pipette recalibration
procedure should be carried out. 
Before starting the recalibration it is necessary to check
whether the following requirements have been fulfilled
during error determination:
• the ambient temperature, and the temperature of the

pipette, tips and water was identical

• the density of the liquid used is close to that of distilled

water

• a balance with appropriate sensitivity has been used.

• mg/μl conversion factor has been taken into account
• the requirements given in sections 3 and 7 have been

fulfilled

If the above conditions are satisfied and the accuracy
error for selected volume given in section 1 exceeds the
permissible value, the pipette recalibration procedure
should be carried out.

The recalibration can be performed within one full
turn of the key to the right or to the left only.

Recalibration conditions:

• Ambient temperature and the temperature of the pipet-

te, tips and liquid should be within the range 20-25°C
and stabilized during weighing within ±0.5°C

• Measurements should be conducted using distilled water
• Balance sensitivity should be suitable for the volume to

be controlled 

10

Recalibration procedure:

• Set the dose volume depending on the pipette volume

according to the following table: 

• Perform 5 aspirations, weigh each one and calculate

the average value of the aspirations 

• Calculate average aspirated volume in μl multiplying

the average aspiration amount [mg] by the distilled
water density coefficient [μl/mg], which depends on
temperature and pressure according to the following
table:

If the average aspirated volume exceeds the permissible
value, the following should be done: 
• Remove the pipetting pushbutton, (Fig. 4A),

Warning: The pipetting pushbutton consists of 2 parts:
the knob (Fig. 1A2) and the pushbutton (Fig. 1A1). After
removal of the pushbutton, both parts are separated.

• Holding the volume adjustment knob to protect it

against rotation, insert the calibration key into the cuts
of the calibration screw, (Fig. 4B),

BPP2

BPP10

BPP20

BPP100

BPP200

BPP1000

BPP5000

BPP10000

0.1 - 2

0.5 - 10

2 - 20

10 - 100

20 - 200

100 - 1000

1000 - 5000

1000 - 10000

0.2

0.5

2

10

20

100

1000

1000

0.176 - 0.224

0.48 - 0.52

1.94 - 2.06

9.84 - 10.16

19.76 - 20.24

98.4-101.6

994 - 1006

975 - 1025

0.06

0.33

0.63

2.50

6.30

25.00

125.00

250.00

Model 

Range 

of the pipette

volumes 

[μl] 

Preset 

volume 

[μl] 

Permissible 

volumes 

[μl] 

Volume change 

Δ

V for full turn 

of the calibration

key [μl]

(24 increments)

Volume checked [μl]

0.1 - 10

10 - 100

> 100

Balance sensitivity [mg]

≤ 

0.001

≤ 

0.01

≤ 

0.1

Temperature [°C]

Pressure [kPa]

95.0

101.3

105.0

20

1.0028

1.0029

1.0029

21

1.0030

1.0031

1.0031

22

1.0032

1.0033

1.0033

23

1.0034

1.0035

1.0036

24

1.0037

1.0038

1.0038

25

1.0039

1.0040

1.0040

Содержание BioPette PLUS BPP2

Страница 1: ......

Страница 2: ...A N F G J I C A1 A2 B P C D E D L A B C D E F...

Страница 3: ...P 5A 5B P 5 4 6 4D 4A 4B 4C 6A 2 1000 l 7 6B 5000 l 10000 l 6C 5000 l 10000 l D D D D D D...

Страница 4: ...US is equipped with a built in tip ejector Fig 1D The construction of the ejector enables the user to adjust the length This allows for proper fit and ejection of nearly every tip type available When...

Страница 5: ...PP2 Red figures at the bottom 1 100 l Increment 0 002 l Pipettes BPP10 BPP20 Red figures at the bottom 1 10 l Increment 0 02 l Pipettes BBPP100 BPP200 Black figures only 1 l Increment 0 2 l Pipettes B...

Страница 6: ...touching the orifice of the tip Dispensing Place the end of the tip against the inside wall of the vessel at an angle of 10 to 40 degrees Press the pushbutton smoothly to the first stop Fig 2C Wait on...

Страница 7: ...s are used for volumes between 0 1 l and 10 l They are used with the BPP2 and the BPP10 models Tips 200 These tips are used for volumes between 2 l and 200 l They are used with the BPP20 BPP100 and th...

Страница 8: ...e 20 25 C and stabilized during weighing within 0 5 C Measurements should be conducted using distilled water Balance sensitivity should be suitable for the volume to be controlled 10 Recalibration pro...

Страница 9: ...f the shaft is cracked or scored Remove the tip ejector Unscrew the shaft nut inspect the shaft and the piston assembly Replace the damaged parts see Section 12 When reasembling the pipette the nut sh...

Страница 10: ...n an autoclave at 121 C for 20 minutes After sterilization the pipette should be dried and cooled to room temperature It is recommended to sterilize the pipettes in autoclave with an initial vacu um a...

Страница 11: ...efore returning a pipette to a Labnet representative please ensure that the pipette is completely free of any contamination chemical microbial or radioactive Contact Labnet at 1 732 417 0700 fax 1 732...