|        LABNET 311DS SHAKING INCUBATOR

 7.0  Calibration

The Labnet 311DS Shaking Incubator is calibrated at 37°C at the factory. The unit can be recalibrated after the chamber temperature 
has stabilized at the setpoint  for several hours. Suspend a certified reference thermometer in the chamber. Compare the units display 
to the reference thermometer. If there is an unacceptable difference, put the controller into calibration mode by pressing both the 
Up and Down arrows at the same time until the two outside decimal points begin to flash. While the decimal points are flashing, the 
display can be calibrated to match the reference thermometer by pressing the Up or Down arrows until the display reads the correct 
value. Allow the shaker incubator to stabilize again, and recalibrate if necessary.

 8.0  Operation

As advised in the installation section, it is very important that the unit be level and placed on a very solid surface. When loading the 
platforms, it is also important to load these in a balanced manner or the unit can vibrate and possibly walk. Shaking unbalanced loads 
for extensive periods may also damage the shaker.

Before powering on the unit, you should first set the mechanical Safety Thermostat on the back of the unit to a temperature just 
above your intended operating temperature or at the maximum (setting of 10).

After you first power on the unit you should notice the shaker display first showing dashes, then the power input line frequency 
(F60 or F50 for 60 Hz or 50 Hz), and the temperature controller display showing some random numbers and dots blinking.

The shaker display and temperature display should show the last time that was set (or Hld) for continuous operation, and the 
temperature display showing the chamber temperature.

You may now set the desired operating temperature, shaking speed, and shaking time (see Controls section). It is recommended that 
you test all loads to be shaken at a slow speed under observation and then gradually raise the speed to ensure the load and unit are 
stable. The unit should also be allowed to equilibrate to the set temperature before using.

Once the temperature, shaking speed, and shaking time are set, use the Stop/Start pad to initiate shaking. When shaking in the timed 
mode, End will be displayed on the shaker display when the shaking is done. In timed mode Hld (Hold) will be displayed on the shaker 
display. The user may switch the shaker display to show the shaking speed by pressing the Push to Shift button.

The user can observe the load being shaken through the glass window and can also turn on the chamber light for additional 
visibility. The chamber light should not be left on for an extended period as it adds heat to the chamber and may cause the chamber 
temperature to rise above the setpoint .

The user may open the chamber door at any time to gain access to the chamber. If the shaker is running when the door is open, the 
shaker power will be immediately cut causing the shaker to slow and stop. The shaker display will show Opn (Open). Shaking will 
automatically resume when the door is closed if there is still time remaining on the shaker run. 

NOTE:

 The clock continues to count 

down when the door is open even though the shaker has stopped.

 9.0  Cleaning and Maintenance

NOTE:

 Be sure to disconnect the power cord before cleaning or decontaminating the oven.

No routine maintenance is required for the electrical or mechanical components of the unit. The incubator exterior, interior, and 
shaking platform should be wiped down periodically with a soft damp cloth with mild soap. 

NOTE:

 Do not use chlorine-based bleach 

or abrasives. Any spills in the incubator and/or on the shaking platforms should be cleaned up immediately. 

Содержание 311DS

Страница 1: ...Labnet 311DS Digital Shaking Incubator Instruction Manual Catalog Numbers I 5311 DS I 5311 DS 230V...

Страница 2: ...ormation 1 3 0 Specifications 2 4 0 Unpacking 2 5 0 Installation 2 6 0 Controls 3 7 0 Calibration 4 8 0 Operation 4 9 0 Cleaning and Maintenance 4 10 0 Troubleshooting and Service 5 11 0 Accessories 5...

Страница 3: ...vey information to the user and to call the users attention to important safety precautions and guides for using this equipment Indicates a potential risk and alerts you to proceed with caution Indica...

Страница 4: ...311DS Shaking Incubator examine the carton and unit for damages If shipping damage has occurred a claim must be filed with the carrier The carrier is responsible for correcting shipping damages Save...

Страница 5: ...s and starts the shaking mechanism and causes the RUN lamp to be illuminated 6 Chamber light switch Turns the chamber light On Off when the power is On 7 Temperature Controller and Temperature Set The...

Страница 6: ...ure You may now set the desired operating temperature shaking speed and shaking time see Controls section It is recommended that you test all loads to be shaken at a slow speed under observation and t...

Страница 7: ...and reference thermometer do not match Be sure the unit has been allowed to stabilize for 1 hour Thermometer should be suspended in the chamber and not touching any surface Only certified reference th...

Страница 8: ...the guidelines specified in the supplied instruction manual This warranty does not cover damage caused by accident neglect misuse improper service natural forces or other causes not arising from defe...

Отзывы: