LOKASET R
| Installation and Operation Instructions | D18067Ae
6
Figure 1.
3. Lokaset R – Installation
The Lokaset R central unit (Figure 1) is installed in a dry indoor
location. A good installation location enables easy reading and
control of the device. Do not cover the central unit as this may
interfere with signal reception from the transmitter. The Lokaset
R central unit is mounted on a wall with screws through the
holes at the bottom of the device. The lid of the housing must
be opened for installation. The device is plugged to the 230 V
mains via the power cord.
Operation of the probe and the transmitter should be tested
during installation in order to ensure that the radio signal can
travel freely. Activate the transmitter test mode (see Service
instructions, p. 9). Immerse the measuring head momentarily in
water. If the central unit is operational, the blue light (SIGNAL)
should turn off every 8 seconds and the alarm should be
signalled within 8 seconds.
Installation of the Lokaset R transmitter is illustrated on the
adjacent page. The transmitter is installed on the cover (1) of
the tank’s outlet pipe using the provided plate screws (2);
4 pcs, 2.9–3 x 25 mm.
Place the rubber gasket (3) between the transmitter and the cover.
The cover of the transmitter must be opened for installation. Make a
lead-through for the cable (4) or for the entire probe in the cover of
the outlet pipe.