4 | System Requirements
This section discusses Access system specifications, including:
l
l
Computer Specifications And Supported Operating System
l
l
Facility Requirements
Power Dual Robot System
l
Two Single-Phase 230VAC, 50/60Hz, 30A drops (regions with 200 - 230V Native Power)
l
Two 3-Phase 5-wire 208VAC, 50/60Hz, 30A drops (regions with 100 - 120V Native Power)
Air
l
Standard:
l
Oil-free (max oil content of 0.01ppm) compressed dry air (CDA) at 620 kPa to 690 kPa (90 PSI to 100 PSI) and flow rate
of 180Lpm/6.3CFM
Vacuum
l
200 Torr, flow rate 70 Lpm.
Laboratory Floor Requirements
Note:
Before purchasing an Access system, a Labcyte Representative will come on site to survey the floor to ensure it
meets the following specifications required for installation when a system has docking capabilities.
To meet docking repeatability specifications, the floor levelness must fall within the range of the following specifications and
standards:
l
ASTM E 1155 Standard
l
FF35/FL25 to FF50/FL30
Laboratory Environment Requirement
l
Ambient temperature: 21°C ± 2°C / relative humidity 10-80% (non-condensing).
Facilities HVAC input requirements for low %RH option
l
Temperature 20°C ± 1°C / relative humidity 10% - 30%
l
Input airflow of 300 - 500CFM
l
Exhaust pressure of -50 Pascals
Building Management System (BMS)
l
Three 24V rated dry contacts for UPS, generator and system environment
Computer Specifications And Supported Operating System
Listed below are the model and specification requirements for the Access DRS.
The Access computer specifications are:
l
Intel® Core™ i7 -6500U, 2.5GHz
l
Memory 16GB (up to 32GB possible)
l
Storage 1x M.2 PCIe/SATA + 1x 2.5" SATA
l
Disk Samsung 960PRO 512GB PCIe NVMe M.2 Internal SSD (MZ-V6P512BW)
l
Graphics Intel® HD Graphics 520
LABCYTE INC.
USER GUIDE|Access™ Dual Robot System
CHAPTER 4 | System Requirements
21
Содержание Access Dual Robot
Страница 25: ...Figure 13 Device Module 1 Right Top View Device Module 1 CHAPTER 5 System Dimensions 25 ...
Страница 27: ...Figure 15 Device Module 2 Top View Device Module 2 CHAPTER 5 System Dimensions 27 ...
Страница 29: ...Figure 17 Echo Module Top View Echo Module CHAPTER 5 System Dimensions 29 ...
Страница 31: ...Figure 19 Storage Module AmbiStore Top View AmbiStore Storage Module CHAPTER 5 System Dimensions 31 ...
Страница 33: ...Figure 21 Storage Module MicroServe Top View MicroServe Storage Module CHAPTER 5 System Dimensions 33 ...
Страница 35: ...Figure 23 Storage Module Cytomat Top View Cytomat Storage Module CHAPTER 5 System Dimensions 35 ...
Страница 71: ...Figure 70 Sample Error Screen Robot Device Recovering From An Emergency Shutdown CHAPTER 7 Basic Operations 71 ...