54
Calibration - Offsets
Adjust the sensor readings to match a known standard(s).
Current Reading (uncorrected)
= raw sensor reading with no offset applied.
Offset
= Value that will be added to the Current Reading (uncorrected). The
adjusted reading (uncor offset) is the value reported on all screens
outside of the Sensor Calibration screen.
1. To adjust the calibration offset, touch
Offset
field of the desired sensor. A pop-up
will display, which allows an offset value to be adjusted for the sensor.
2. Using a known standard as reference, change the offset to match the value
reported by the standard. For example, if the
Sump / Water Temp
reports an
uncorrected reading of 35
o
C, and the standard reference reports 36
o
C, change
the offset to +1
o
C.
When the
Offset
value is changed, the current date will be saved as the
Last
Cal. Date
Offset limits:
Temperature = +/- 5
o
C
Conductivity = +/- 100
µ
S
Fill Level = +/- 5%
Figure 6-16