![Labconco RapidStill II Скачать руководство пользователя страница 25](http://html1.mh-extra.com/html/labconco/rapidstill-ii/rapidstill-ii_user-manual_3162924025.webp)
Product Service 1-800-522-7658
22
C
C
h
h
a
a
p
p
t
t
e
e
r
r
6
6
:
:
T
T
r
r
o
o
u
u
b
b
l
l
e
e
s
s
h
h
o
o
o
o
t
t
i
i
n
n
g
g
PROBLEM
CAUSES
CORRECTIVE ACTION
Unit will not operate
at all
Tripped circuit breaker
Check facility’s circuit breaker
Blown fuse
Replace fuse
Power cord not plugged
in
Plug in cord
Boiler doesn’t fill
‘Boiler Heater’ switch
is on
Turn off boiler heater before
filling
Cooling water off or
flow rate insufficient
Check flow rate of condenser
cooling water
Water fill valve plugged
or defective
Clean inlet screen of filter;
replace valve
‘Boiler Water’ switch
defective
Replace switch
Heating element
doesn’t work
Heating element is
defective
Replace heating element
‘Boiler Heater’ switch
is defective
Replace switch