Quickstart
How do I recalibrate the
blower speed
filter life gauge
airflow sensor reading
of the Logic+?
1.
Press the
Menu
button.
2.
Press the
▼
button until the
Services
option is highlighted (it will
turn yellow).
3.
Press
OK/Mute
to enter the Services menu screen.
4.
Press the
▼
button until the
Calibration
option is highlighted (it
will turn white).
5.
Press
OK/Mute
to start the calibration procedure.
6.
You will see a
warning screen
alerting you that you are about to
alter the BSC’s settings.
7.
Press
OK/Mute
to input the password.
8.
When requested for the password press
Light
,
UV light
,
Timer
,
Timer
then
OK/Mute
.
9.
If the password is properly entered, the first certification screen
will display the blower motor’s PWM setting.
10.
Press the
▲
or
▼
button to increase or decrease the blower
speed as needed. When the speed is set properly, press the
OK/Mute
button to lock the new blower speed setting, and
continue to the filter life screen.
MAIN MENU
1. MyLogic
TM
2. Settings
3. Service
or
to choose,
OK to select, MENU to escape
or
to choose,
OK to select, MENU to return
Current Settings
Restore Default
Diagnostics
Configuration
Calibration
Service
ATTENTION!
The following screens are used
to alter the operation of the
Logic. They should only be used
by a qualified certifier as part
of the certification process.
Calibration
OK to proceed, MENU to return
PASSWORD:
Calibration
The ECM setting is: 80
The current RPM is: 1050
The average RPM is: 1020
Calibration
Motor Speed Adjustment
or
to change motor speed,
OK to select, MENU to finish