Original instructions 

 
 
 

A

PPENDIX 

E

 

S

ERIAL

 N

UMBER 

T

AG 

D

ESCRIPTION 

 

Serial tag includes standard information with the following changes: 
 
The portion of the tag for electrical information is labeled with the following 
units: Volts AC, Phase, Cycle, AMP. 
 
Each hood will include one of the following:  
 

Volts AC 

Phase 

Cycle 

AMP    x 

115 

50/60 

10X 

115 

50 

10X 

115 

60 

10X 

230 

50/60 

5X 

230 

50 

5X 

230 

60 

5X 

 

Where 

 may be a separate label indicating: Caution, See Manual. 

 
The X above will be one of several possible alpha characters as defined in the 
manual as follows: 
 
X: 

No additional duplexes on hood, noted amperage applies. 

B: 

This fume hood includes 

one 

additional electrical receptacle, individually 

wired to the field wired box and individually rated as 115V, single phase, 
60Hz, 20 Amps.  Each duplex can be wired on a dedicated circuit rated at 
20A or the duplexes may be ganged together on the same circuit with the 
total load of the ganged duplexes not exceeding the 20A electrical service 
they are wired to. 

C: 

This fume hood includes 

two

 additional electrical receptacles, individually 

wired to the field wired box and individually rated as 115V, single phase, 
60Hz, 20 Amps.  Each duplex can be wired on a dedicated circuit rated at 
20A or the duplexes may be ganged together on the same circuit with the 
total load of the ganged duplexes not exceeding the 20A electrical service 
they are wired to. 

Содержание Protector XL 1110 Series

Страница 1: ...Original instructions Protector XL Laboratory Fume Hoods Models 1110 Series 1111 Series 1112 Series 1113 Series 1114 Series 1115 Series 1116 Series 1117 Series 1118 Series...

Страница 2: ...e hours of 7 30 a m and 5 30 p m Central Standard Time Part 9410700 Rev L ECO N065 Warranty Labconco Corporation provides a warranty to the original buyer for the repair or replacement of parts and re...

Страница 3: ...equirements 8 CHAPTER 3 GETTING STARTED 9 Unpacking Your Fume Hood 10 Removing the Shipping Skid 10 Sash Weight Release 11 Installing the Hood on a Supporting Structure Work Surface 11 Connecting to t...

Страница 4: ...e Hood 34 Installing Additional Service Fixtures 35 Installing Guardian Airflow Monitors 35 Distillation Grids Field Installation 35 Sash Stop Kit Field Installation 36 Upper Bypass Field Installation...

Страница 5: ...a laboratory and effectively contain toxic noxious or other harmful materials when properly installed The Protector XL offers many unique features to enhance safety performance and visibility To take...

Страница 6: ...s you should take when using the fume hood Chapter 5 Using Your Protector XL discusses the basic operation of your fume hood Information on how to prepare use and shut down your Protector Hood are inc...

Страница 7: ...information following an exclamation icon may result in injury to the user or permanent damage to fume hood Les informations critiques sont pr sent es en gras dans les paragraphes qui sont pr c d s p...

Страница 8: ...h hazard PRUDENCE Consulter le Manuel Ce symbole sur la hotte indique la possibilit d un risque de pincement Your Next Step If your Fume Hood needs to be installed proceed to Chapter 2 Prerequisites t...

Страница 9: ...of electrical power must be located near the installation site Carefully read this chapter to learn the requirements for your installation site The location requirements The support requirements The...

Страница 10: ...or mobile bench ALL Protector Hood installations must be permanent and stationary The supporting structure usually consists of a base cabinet and chemically resistant work surface NE PAS installer la...

Страница 11: ...ype Face Velocity fpm Airflow Volumetric Rate CFM Airflow Volumetric Rate CFM Sash at Sash at 62 5 Static Pressure inches of water Static Pressure inches of water Full open open 4 5 6 8 4 5 6 8 at 28...

Страница 12: ...c blage interne est termin la bo te de jonction point de c blage unique pour le raccordement par un lectricien qualifi Le bouton du ventilateur et fils de l interrupteur de lumi re sont galement mis...

Страница 13: ...wrenches ratchets sockets a nut driver set a flat blade screwdriver a Phillips screwdriver and a carpenter level to complete the instructions in the chapter The Protector XL Hood models weigh between...

Страница 14: ...ard The upper baffle is 16 5 tall the middle baffle is 23 3 tall and the lower perforated baffle is 17 3 tall See Figure 1 1 Figure 3 1 and Figure B 1 that shows proper XL baffle installation Do not r...

Страница 15: ...when lifting or moving the unit La Hotte Protecteur est lourd Prudence en soulevant ou en d pla ant l objet When installing the Protector Fume Hood onto a chemically resistant work surface or benchto...

Страница 16: ...or base cabinet 4 Place the hood on top of the work surface and cross support The work surface should be smooth and durable such as a chemically resistant epoxy resin The surface should be nonporous a...

Страница 17: ...g your blower for the exhaust volume and total system static pressure loss La connexion d chappement devrait tre install e par un professionnel de CVC agr e La connexion d aspiration sur votre hotte a...

Страница 18: ...ification The number of circuits varies depending on the model All of the electrical connections are terminated at the field wiring terminal box for hook up by a qualified electrician We recommend eac...

Страница 19: ...nduit to protect the wiring into the hood The grounding connection shall not be made to the terminal box cover Tout le c blage lectrique pour la hotte aspirante devrait tre ex cut par un lectricien ag...

Страница 20: ...e de fuite Resserrer les installations l g rement si n cessaires CAUTION Do not use oxygen with any standard service fixture Contact Labconco Customer Service for oxygen fixture information PRUDENCE N...

Страница 21: ...you the flexibility to change the airflow at the sash opening of your hood To determine the actual face velocity at the sash opening airflow velocity readings will need to be taken This should be don...

Страница 22: ...ntrol fluctuations in face velocity which occur as the sash is closed 1 Unique sash provides maximum visibility of 37 5 high while conserving energy by limiting sash travel to 28 Vertical rising sash...

Страница 23: ...s provide maximum visibility and the flexibility to add services after installation 11 All hoods are factory prepared for up to 8 service fixtures 12 Duplex electrical receptacles are mounted on the r...

Страница 24: ...Chapter 4 Performance Features Safety Precautions Product Service 1 800 522 7658 20 Original instructions 17 7 14 18 10 8 9 2 15 1 12 5 11 20 16 6 Figure 4 1 22 4 21 23...

Страница 25: ...la hotte de laboratoire ait t r gl e pour maintenir la s curit optimale de l op rateur la prudence devrait toujours tre utilis e en travaillant sous la hotte Avant utiliser la hotte le contr le pour s...

Страница 26: ...as possible It is best to keep all chemicals and apparatus 6 inside the front of the hood Toujours travailler aussi loin que possible de la hotte Il est recommand de garder tous les produits chimique...

Страница 27: ...te de Laboratoire Protecteur devrait tre certifi par un technicien de certification qualifi avant qu elle soit utilis e au d but La hotte devrait tre re certifi e quand elle est r install e entretenue...

Страница 28: ...surfaces of the motor can become hot and could cause burns Ne pas toucher le moteur de soufflerie Les surfaces du moteur peuvent devenir chaudes et pourrait causer des br lures Do not position the fu...

Страница 29: ...ric Intelli Sash are available Operating the A Style Combination Sash Some other hood models have additional energy saving sashes called A Style Combination Sashes in place of vertical rising sashes T...

Страница 30: ...sure is an excellent way to ensure hazardous fumes in the fume hood do not enter the laboratory space inadvertently Turn on the light Slowly raise the sash and verify the fume hood has proper airflow...

Страница 31: ...or two minutes undisturbed to purge airborne contaminants from the work area before closing the sash Unloading Materials and Equipment Objects in contact with contaminated material should be surface d...

Страница 32: ...operations require a step ladder always use proper safety and consult your safety officer If performing maintenance on any service lines always shut off the supply first Some removable components may...

Страница 33: ...ions d entretien qui exigent des techniciens qualifi s Routine Maintenance Schedule Weekly Using ordinary dish soap to clean the surface inside of the fume hood and the work surface Using an appropria...

Страница 34: ...pirante par un technicien agr en certification Voir au Chapitre 3 Certifier la Hotte Aspirante Protecteur All monthly activities Toutes les activit s mensuelles Biannually The sash assembly should be...

Страница 35: ...your individual requirements These include the addition of work surfaces service fixtures air monitor distillation grids electrical duplex outlets ceiling enclosures and rear panels Installing Work S...

Страница 36: ...Fixed Height Ceiling Enclosure Kit Catalog Number Use this key to configure the seven digit catalog number to order your 3 Sided Fixed Height Ceiling Enclosure Kit For example 9550511 is a 3 Sided Cei...

Страница 37: ...EP 2 Select the depth based on the model chosen Choose 0 for Premier and 1 for XStream Choose 1 for PVC Choose 0 1 or 2 for the correct depth of the XL Benchtop model Choose 0 or 1 for the correct dep...

Страница 38: ...Protector Fume Hood can be modified to add a rear panel behind the fume hood when the fume hood is placed on an island not available on Pass Through Fume Hood Contact Labconco Customer Service for ord...

Страница 39: ...rdian Airflow Monitors continuously monitor face velocity through the fume hood opening The fume hood right corner post is factory prepared to mount either monitor Contact Labconco Customer Service to...

Страница 40: ...n have one installed in the field by a qualified electrician Contact Labconco Customer Service for ordering information Not acceptable on explosion proof hoods Votre Prot ge capot de fum es peut tre c...

Страница 41: ...oxes or switches Check connection of switches Check connection to control box on top of unit Circuit breakers tripped in building electrical supply Reset circuit breakers Remote blower won t operate b...

Страница 42: ...contamination See Location Requirements Chapter 2 Certifying the Hood Chapter 3 and Safety Precautions Chapter 4 sections of this manual Ref Appendix E Fume hood has improper face velocity Have fume h...

Страница 43: ...tely if damaged Electrical duplex outlets no longer have power Wires not connected or faulty duplex Check wire connection or replace duplex Circuit breakers tripped in building electrical supply Reset...

Страница 44: ...ver Valve Connector OXYGEN LIGHT GREEN 2J 1 9826810 Swivel Gooseneck only GRN 2K 1 9826812 Swivel Gooseneck only WHITE 3 1 9818700 thru 08 Knobs GRAY GRN BLU ORG YEL RED WHT BLK BRN 4A 1 9818800 Hose...

Страница 45: ...5308 Screw Machine 10 24 x 50 Phillips 13A 1 9436500 Eco Foil 3 13B 1 9436501 Eco Foil 4 13C 1 9436502 Eco Foil 5 13D 1 9436503 Eco Foil 6 13E 1 9436504 Eco Foil 7 13F 1 9436505 Eco Foil 8 13G 1 94743...

Страница 46: ...Appendix A Protector XL Components Product Service 1 800 522 7658 42 Original instructions Figure A 1 15 12 10 21 5 6 2 13 7 1 3 11 22 14 16 18 8 4...

Страница 47: ...C L One Duct C L One Duct 48 00 Two Ducts 50 00 Two Ducts 62 00 Two Ducts 48 00 Four Ducts Hood Size DIM A DIM B DIM C DIM D 3ft 24 Depth 36 00 31 7 26 25 24 30 Depth 37 7 30 36 Depth 43 7 36 4ft 24...

Страница 48: ...Transient over voltages according to Installation Categories II Over voltage Categories per IEC 1010 Temporary voltage spikes on the AC input line that may be as high as 1500V for 115V models and 250...

Страница 49: ...Appendix C Protector XL Specifications Product Service 1 800 522 7658 45 Original instructions...

Страница 50: ...Appendix C Protector XL Specifications Product Service 1 800 522 7658 46 Original instructions...

Страница 51: ...ated Protector I S Hoods with limit switches are only available from the factory due to the wiring complexity The integrated Protector I S Hood models with limit switches automatically regulate the In...

Страница 52: ...change automatically with sash position and room controls work independently Room controls are sold separate and not sold by Labconco INTELLIGENT The Integrated Intelli Sense Hood becomes a Smart Hood...

Страница 53: ...and set to activate with sash raised 20 above work surface The upper limit switch can be lowered in 1 increments between 15 and 20 Red wire to common White wire to normally open Intelli Sense Control...

Страница 54: ...Appendix D Protector I S Hood Product Service 1 800 522 7658 50 Original instructions Red wire to common Left side lower limit switch wired for low speed Black wire to normally open...

Страница 55: ...ons Airflow monitor in line sensor air pathway and wiring to monitor Night setback lower switch shown on right Switch is wired to normally open and common terminals and wired to air monitor input I P1...

Страница 56: ...Appendix D Protector I S Hood Product Service 1 800 522 7658 52 Original instructions Front of airflow monitor showing SETBACK when sash is closed...

Страница 57: ...x D Protector I S Hood Product Service 1 800 522 7658 53 Original instructions Back of airflow monitor night setback switch wiring to I P1 for Input 1 Sensor wiring in to air monitor Power in to air m...

Страница 58: ...Appendix D Protector I S Hood Product Service 1 800 522 7658 54 Original instructions 115V Wiring Diagram Automatic Operation 230V Wiring Diagram Automatic Operation...

Страница 59: ...exes on hood noted amperage applies B This fume hood includes one additional electrical receptacle individually wired to the field wired box and individually rated as 115V single phase 60Hz 20 Amps Ea...

Страница 60: ...xes may be ganged together on the same circuit with the total load of the ganged duplexes not exceeding the 10A electrical service they are wired to H This fume hood includes three additional electric...

Страница 61: ...xes may be ganged together on the same circuit with the total load of the ganged duplexes not exceeding the 10A electrical service they are wired to N This fume hood includes four additional electrica...

Страница 62: ...tiers c bl es individuellement et class comme 115V monophas 60 Hz 20 amp res Chaque duplex peut tre branch sur un circuit d di valu 20A ou les duplex peuvent tre group s ensemble sur le m me circuit q...

Страница 63: ...ent et class comme 230V monophas 60 Hz 10 amp res Chaque duplex peut tre branch sur un circuit d di class au 10A ou les duplex peuvent tre group s ensemble sur le m me circuit que la charge totale des...

Страница 64: ...ge totale des duplex coupl es ne d passant pas le service lectrique 10A ils sont c bl s N Cette hotte comprend quatre autres prises lectriques individuellement c bl s la bo te c bl es individuellement...

Страница 65: ...od face velocities between 60 100 lfpm Maximum of 125 lfpm for radioisotope hoods Duct velocities of 1000 2000 fpm for vapors gasses and smoke Stack discharge height 1 3 2 0 x building height Well des...

Страница 66: ...ng Chemicals Laboratory hoods should not be relied on for explosion protection Exhaust air from fume hoods should not be recirculated Services should be external to the hood Canopy hoods only for non...

Страница 67: ...worth Publishers 1985 Miller Brinton M et al Laboratory Safety Principles and Practices American Society for Microbiology Washington D C 1986 NIH Guidelines for the Laboratory Use of Chemical Carcinog...

Страница 68: ...s Nut Driver Set Decorative Panel Removal 1 Remove side panels by removing 10 24 screws and lifting away 2 Remove front panels by lifting up and away Light Removal and Blower Wiring 1 On the top of th...

Страница 69: ...e designed with one continuous top liner Therefore the top support is not required for the 4 through 8 hoods but is used on the 10 12 and 16 hoods 1 Remove the top support by removing the 1 4 20 hardw...

Страница 70: ...doubt about lifting the hood always reassemble the hood where it will be placed Reinstallation Make sure the anti rack shaft does not slide from left to right as this will be noisy and not work effec...

Страница 71: ...this requirement Labconco offers the following step by step procedure to ensure you have the correct information 1 Locate the serial tag for your fume hood usually on the lower right of the sash glass...

Страница 72: ...34 inches A x 0 34 bypass area Bypass area ___________ square inches D sq in b Divide this by 144 to arrive at square feet Bypass area ____________ square feet D sq ft 3 Determining Total Effective O...

Отзывы: