Appendix D: Serial Number Tag Description
Product Service 1-800-522-7658
128
Original instructions
D:
This fume hood includes
three
additional electrical receptacles,
individually wired to the field wired box and individually rated as 115V,
single phase, 60Hz, 20 Amps. Each duplex can be wired on a dedicated
circuit rated at 20A or the duplexes may be ganged together on the same
circuit with the total load of the ganged duplexes not exceeding the 20A
electrical service they are wired to.
E:
This fume hood includes
four
additional electrical receptacles,
individually wired to the field wired box and individually rated as 115V,
single phase, 60Hz, 20 Amps. Each duplex can be wired on a dedicated
circuit rated at 20A or the duplexes may be ganged together on the same
circuit with the total load of the ganged duplexes not exceeding the 20A
electrical service they are wired to.
F:
This fume hood includes
one
additional electrical receptacle, individually
wired to the field wired box and individually rated as 230V, single phase,
60Hz
, 10 Amps. Each duplex can be wired on a dedicated circuit rated at
10A or the duplexes may be ganged together on the same circuit with the
total load of the ganged duplexes not exceeding the 10A electrical service
they are wired to.
G:
This fume hood includes
two
additional electrical receptacles, individually
wired to the field wired box and individually rated as 230V, single phase,
60Hz
, 10 Amps. Each duplex can be wired on a dedicated circuit rated at
10A or the duplexes may be ganged together on the same circuit with the
total load of the ganged duplexes not exceeding the 10A electrical service
they are wired to.
H:
This fume hood includes
three
additional electrical receptacles,
individually wired to the field wired box and individually rated as 230V,
single phase,
60Hz
, 10 Amps. Each duplex can be wired on a dedicated
circuit rated at 10A or the duplexes may be ganged together on the same
circuit with the total load of the ganged duplexes not exceeding the 10A
electrical service they are wired to.
J:
This fume hood includes
four
additional electrical receptacles,
individually wired to the field wired box and individually rated as 230V,
single phase,
60Hz
, 10 Amps. Each duplex can be wired on a dedicated
circuit rated at 10A or the duplexes may be ganged together on the same
circuit with the total load of the ganged duplexes not exceeding the 10A
electrical service they are wired to.
K:
This fume hood includes
one
additional electrical receptacle, individually
wired to the field wired box and individually rated as 230V, single phase,
50Hz
, 10 Amps. Each duplex can be wired on a dedicated circuit rated at
10A or the duplexes may be ganged together on the same circuit with the
total load of the ganged duplexes not exceeding the 10A electrical service
they are wired to.
L:
This fume hood includes
two
additional electrical receptacles, individually
wired to the field wired box and individually rated as 230V, single phase,
50Hz
, 10 Amps. Each duplex can be wired on a dedicated circuit rated at
10A or the duplexes may be ganged together on the same circuit with the
total load of the ganged duplexes not exceeding the 10A electrical service
they are wired to.
Содержание Protector Airo 184 Series
Страница 73: ...Chapter 6 Maintaining the Protector Filtered Hood Product Service 1 800 522 7658 68 Original instructions...
Страница 74: ...Chapter 6 Maintaining the Protector Filtered Hood Product Service 1 800 522 7658 69 Original instructions...
Страница 77: ...Chapter 6 Maintaining the Protector Filtered Hood Product Service 1 800 522 7658 72 Original instructions...
Страница 94: ...Chapter 6 Maintaining the Protector Filtered Hood Product Service 1 800 522 7658 89 Original instructions...
Страница 96: ...Chapter 6 Maintaining the Protector Filtered Hood Product Service 1 800 522 7658 91 Original instructions...
Страница 131: ...Appendix C Protector Filtered Hood Specifications Product Service 1 800 522 7658 126 Original instructions...
Страница 141: ...Appendix E gGuard Product Service 1 800 522 7658 136 Original instructions Explanation of gGuard...
Страница 161: ...Appendix H Chemical Guide Product Service 1 800 522 7658 156 Original instructions...
Страница 162: ...Appendix H Chemical Guide Product Service 1 800 522 7658 157 Original instructions...
Страница 163: ...Appendix H Chemical Guide Product Service 1 800 522 7658 158 Original instructions...
Страница 164: ...Appendix H Chemical Guide Product Service 1 800 522 7658 159 Original instructions...
Страница 165: ...Appendix H Chemical Guide Product Service 1 800 522 7658 160 Original instructions...
Страница 166: ...Appendix H Chemical Guide Product Service 1 800 522 7658 161 Original instructions...
Страница 167: ...Appendix H Chemical Guide Product Service 1 800 522 7658 162 Original instructions...
Страница 168: ...Appendix H Chemical Guide Product Service 1 800 522 7658 163 Original instructions...
Страница 169: ...Appendix H Chemical Guide Product Service 1 800 522 7658 164 Original instructions...
Страница 170: ...Appendix H Chemical Guide Product Service 1 800 522 7658 165 Original instructions...
Страница 171: ...Appendix H Chemical Guide Product Service 1 800 522 7658 166 Original instructions...
Страница 172: ...Appendix H Chemical Guide Product Service 1 800 522 7658 167 Original instructions...
Страница 173: ...Appendix H Chemical Guide Product Service 1 800 522 7658 168 Original instructions...
Страница 174: ...Appendix H Chemical Guide Product Service 1 800 522 7658 169 Original instructions...
Страница 175: ...Appendix H Chemical Guide Product Service 1 800 522 7658 170 Original instructions...
Страница 191: ...Appendix K Small 54 2 Height Airo Dimensions Product Service 1 800 522 7658 186 Original instructions Figure K 1...
Страница 192: ...Appendix K Small 54 2 Height Airo Dimensions Product Service 1 800 522 7658 187 Original instructions Figure K 2...
Страница 193: ...Appendix K Small 54 2 Height Airo Dimensions Product Service 1 800 522 7658 188 Original instructions Figure K 3...