![Labconco Guardian Jr. 4865000 Скачать руководство пользователя страница 11](http://html1.mh-extra.com/html/labconco/guardian-jr-4865000/guardian-jr-4865000_user-manual_667621011.webp)
Chapter 2: Installation
Product Service 1-800-522-7658
7
Protector Hoods
115 Volt: Locate the black wire, which connects from the incoming AC
power (in the junction box) to the hood light switch. Disconnect this black
wire from the light switch terminal and connect the solid black wire in
wiring harness #48610. This black wire will have two extra terminals
attached – one of these terminals should be reconnected to the light switch
terminal. The other terminal should be connected to the unused blower
switch terminal located near the center of the blower switch. Next,
connect the black/white wire, which is a part of harness #48610 to the
remaining blower switch terminal. The final wiring step involves the
white wire, which is part of #48610. Locate the white wire that connects
to the fluorescent lamps and find the point where it connects, via
removable terminals, to the white wire from the junction box. Disconnect
the removable terminal and use the white wire (part of #48610) to
reconnect the white wires. The other attached white wire provides a
connection to the air monitor power supply module.
230 Volt: Locate the black wire, which is connected to the hood blower
switch. Disconnect this black wire from the blower switch terminal and
connect to the blue wire of #48611 using the attached terminal. Reconnect
the remaining terminal of the blue wire (part of #48611) to the original
blower switch terminal. Locate the white wire on the blower switch,
which is adjacent to the blue wire just connected above. Disconnect this
white wire from the blower switch terminal and connect to the brown wire
in wiring harness #48611. Reconnect the remaining terminal on this
brown wire to the blower switch terminal where the white wire was
originally removed.
8 Foot, 230 Volt Hoods with External Blower
230 Volt: Locate the blue wire, which is connected to the blower switch.
Disconnect this blue wire from the blower switch terminal and connect to
the blue wire in wiring harness #48611 using the attached terminal.
Reconnect the remaining terminal of the blue wire (part #48611) to the
original blower switch terminal. Locate the brown wire on the blower
switch, which is adjacent to the blue wire just connected above.
Disconnect this brown wire form the blower switch terminal and connect
to the brown wire in wiring harness #48611. Reconnect the remaining
terminal on this brown wire to the blower switch terminal where the
brown wire was originally removed.
Basic 47 and 70 Fume Hoods
115 Volt: Locate the black wire, which connects from the incoming AC
power (in junction box) to the hood light switch. Disconnect this black
wire form the light switch terminal and connect to the solid black wire,
which is part of #48610. This black wire will now have two extra
terminals attached – one of these terminals should be reconnected to the
light switch terminal. The other terminal should be connected to the
Содержание Guardian Jr. 4865000
Страница 4: ......
Страница 14: ...Chapter 2 Installation Product Service 1 800 522 7658 10 ...
Страница 15: ...Chapter 2 Installation Product Service 1 800 522 7658 11 ...
Страница 18: ...Chapter 4 Electrical Wiring Diagrams Product Service 1 800 522 7658 14 ...
Страница 19: ...Chapter 4 Electrical Wiring Diagrams Product Service 1 800 522 7658 15 ...
Страница 20: ...Chapter 4 Electrical Wiring Diagrams Product Service 1 800 522 7658 16 ...
Страница 21: ...Chapter 4 Electrical Wiring Diagrams Product Service 1 800 522 7658 17 ...
Страница 22: ...Chapter 4 Electrical Wiring Diagrams Product Service 1 800 522 7658 18 ...
Страница 23: ...Chapter 4 Electrical Wiring Diagrams Product Service 1 800 522 7658 19 ...
Страница 24: ...Chapter 4 Electrical Wiring Diagrams Product Service 1 800 522 7658 20 ...
Страница 25: ...Chapter 4 Electrical Wiring Diagrams Product Service 1 800 522 7658 21 ...