Chapter 6: Accessories for Your CentriVap
Product Service 1-800-522-7658
23
Be sure to use the proper cartridge for your application. The moisture cartridge is
used to trap water vapor. The acid cartridge is used to trap acid vapors. The
solvent cartridge is used to trap solvent vapors and the radiochemical cartridge is
used to trap radioactive waste.
When the media in the insert has changed color, discard the insert and replace it
with a new insert. For the radiochemical trap insert, no color indicator exists
therefore it should be discarded after each use. Use a Geiger counter to monitor
the pump exhaust.
Warning: This radiochemical cartridge does not meet NRC filter design
recommendations.
After operating, properly dispose of all hazardous materials in compliance
with all applicable codes. Labconco is not responsible for improper disposal
of any materials.
Installing a Vacuum Gauge
A user-supplied vacuum gauge may be attached to the CentriVap to monitor the
vacuum level. Attach the vacuum gauge to the barb end of the Quick Disconnect
Coupling fitting that was supplied with the CentriVap using a length of suitable
rubber hose. The Quick Disconnect Coupling fitting can then be pushed into the
mating connector on the left side of the CentriVap located towards the rear. The
Quick Disconnect Coupling fitting can be removed from the mating connector by
pressing on the tab on the connector and then pulling out the Quick Disconnect
Coupling fitting. Always remove the Quick Disconnect Coupling from the
CentriVap when a vacuum gauge is not attached to it.
Содержание CentriVap DNA Systems
Страница 1: ...CentriVap DNA Systems Models 7970010 7970011 7970030 7970035 7970037...
Страница 32: ...Appendix A CentriVap Components Product Service 1 800 522 7658 28...
Страница 33: ...Product Service 1 800 522 7658 29 Appendix B CentriVap Dimensions...
Страница 35: ...Appendix C CentriVap Specifications Product Service 1 800 522 7658 31...