45
Product Service: Domestic 1-800-522-7658, International 816-333-8811
Chapter 7: Troubleshooting
The next test allows you to check the pump and the
connections from the pump to the vacuum tube.
1. Remove the screw from the center of the
vacuum tube cover and slide the cover off.
2. Take a 3 or 4 inch piece of vacuum hose and
place it over the vacuum tube. Stopper and
clamp the vacuum hose to the tube. Restart the
system and check your vacuum indication.
Insufficient vacuum indicates a bad pump or a
leak in the connections from the pump to the
bottom of the chamber.
Vacuum Tube Cover
Stopper
Vacuum Tube
Collector Coil
Clamp
4" piece of rubber
tubing
Checking the connections to the pump and the vacuum tube
Содержание 7753000
Страница 2: ...Product designs are subject to change without notice 2002 Labconco Corporation Printed in U S A ...
Страница 4: ......
Страница 18: ...12 Product Service Domestic 1 800 522 7658 International 816 333 8811 Chapter 2 Prerequisites ...
Страница 38: ...32 Product Service Domestic 1 800 522 7658 International 816 333 8811 Chapter 4 Using Your Freeze Dryer ...
Страница 57: ...51 Product Service Domestic 1 800 522 7658 International 816 333 8811 Appendix A Freeze Dryer Components ...
Страница 58: ...52 Product Service Domestic 1 800 522 7658 International 816 333 8811 Appendix A Freeze Dryer Components ...
Страница 60: ...54 Product Service Domestic 1 800 522 7658 International 816 333 8811 Appendix B Freeze Dryer Dimensions ...
Страница 74: ...68 Product Service Domestic 1 800 522 7658 International 816 333 8811 Appendix D Freeze Dryer Accessories ...
Страница 75: ...69 Product Service Domestic 1 800 522 7658 International 816 333 8811 ...