INSTALLATION 

The most effective installation is one in which the back is mounted against a wall while the front is 
supported by 3/8 inch threaded rod mounted to the ceiling.  Wire or cable may be substituted for 
the 3/8 inch threaded rod.  Please see Figure 1 for recommended installation technique. 
 

 

 

Figure 1 

 

Hood Efficiency 

 
The Canopy Hood can be extremely effective when installed correctly.  The effectiveness of the 
Canopy Hood is dependent directly upon the proximity of the work area to the hood face.  The 
face velocity and static pressure expressed in the Performance section is based on a theoretical 
hood face at 1 foot below the hood.  If the height of the working apparatus is below the theoretical 
1-foot face, then efficiencies will drop off.  Conversely, the hood will become more efficient as the 
height of the working apparatus enters inside the 1-foot face. 
 
Therefore, the distance between the work surface and the hood is extremely important.  Even when 
used as a multipurpose installation, it is important to be within or near the 1-foot face for 
maximum efficiency for all uses.  Refer to Figure 2. 
 

Содержание 4799000

Страница 1: ...CANOPY HOOD INSTRUCTION MANUAL Models 4799000 and 7099000 Product designs are subject to change without notice 1994 2007 Labconco Corporation 4809400 Revision C ECO E427 Printed in U S A ...

Страница 2: ...TABLE OF CONTENTS Introduction Components Shipped 1 General Description 2 Performance 2 Installation 3 Safety Precautions 5 Dimensions 6 Warranty Contacting Labconco 7 ...

Страница 3: ...T NOTICE This manual should be read carefully by all the end users in order to become familiar with the operation of the Canopy Hood Recommendations are made within the manual to help you obtain maximum performance and life from your product We have included sections on performance initial set up and safety precautions to provide you with all the tools necessary to achieve maximum performance If y...

Страница 4: ...it Do not discard the carton or packing material until all components have been checked against the following component list and the equipment has been installed and tested As shipped the carton should contain the following Part Number Description 47990 4 Foot Canopy Hood 70990 6 Foot Canopy Hood 48094 Instruction Manual ...

Страница 5: ... All component parts of the Canopy Hood are epoxy coated except the baffle support brackets which are stainless steel The baffle design enhances perimeter air velocities and capture efficiency Static pressure loss through the hood shall not exceed 64 inches at 907 CFM and 1 25 inches at 1 360 CFM for the 4 foot unit Static pressure loss through the 6 foot unit shall not exceed 90 inches at 1 289 C...

Страница 6: ... upon the proximity of the work area to the hood face The face velocity and static pressure expressed in the Performance section is based on a theoretical hood face at 1 foot below the hood If the height of the working apparatus is below the theoretical 1 foot face then efficiencies will drop off Conversely the hood will become more efficient as the height of the working apparatus enters inside th...

Страница 7: ...INSTALLATION 4 Figure 2 ...

Страница 8: ...SAFETY PRECAUTIONS 5 Note Canopy Hoods are not provided with explosion proof electrical components Flammable or explosive atmospheres should not be used in conjunction with this hood ...

Страница 9: ...DIMENSIONS 6 ...

Страница 10: ...es areas of defective material and workmanship provided such defect results from normal and proper use of the equipment The warranty for all Labconco products will expire one year from date of installation or two years from date of shipment from Labconco whichever is sooner except the following Purifier Delta Series Biological Safety Cabinets and PuriCare Lab Animal Research Stations carry a three...

Отзывы: