Your 

Labcold

 pharmacy and vaccine fridge is fitted as standard with a door lock with 2 keys 

unless the digital lock has been specified. To lock, push the key in gently and twist 90⁰

 

to the 

left, to unlock twist 90⁰

 

to the right. The key can be safely removed from both the locked and 

unlocked positions.

Keys should not be left in the lock as they protrude from the fridge. They could cause a 

health and safety hazard or could be knocked, which may damage the lock. 

The display on your fridge is designed to display error codes. If you see one it is because the 

controller, which performs continuous diagnostic system checks, has detected a problem. 

Most require immediate action. 

batt

 - this means that the batteries 

need changing. It will only display if 

batteries are fitted and you can still use 

the fridge safely

Poff 

when this displays it means 

there is or has been a power failure. If 

batteries are fitted this will display even 

if the mains power is off

Start Error 

when this displays it 

means there is a total system failure. Do 

not use the fridge, remove contents. 

Please contact your retailer

NTC 

when this displays it means that 

the temperature probe has failed. Do 

not use the fridge, remove contents. 

Please contact your retailer

If, when the fridge is first plugged in the display shows 

test

 or it continually displays the first 

two digits of the model number ie 02

 

(with no decimal point) or the display does not change 

to an alarm state or show a temperature value after a few minutes, there is a system failure.  Do 

not use the fridge, contact your retailer.

If the display is blank it means either that the controller is not working correctly or the fridge 

is not getting power. You can check this. If the light comes on when the door is opened and 

you can hear the fridge working and it feels cold inside it means the display may be faulty. If 

the light does not come on and the fridge does not feel cold do not use the fridge, please 

remove contents and contact your retailer.

Error Codes

Door Lock

COLUMN

CODE

A

Time

Time of reading

B

T1

Temperature Probe 1

C

T2

Temperature Probe 2 ( if fitted)

D

1(on) or 0(off)

Door Open

E

1(on) or 0(off)

Compressor Running

F

1(on) or 0(off)

Fan Running

G

1(on) or 0(off)

Light on

H

1(on) or 0(off)

High alarm

I

1(on) or 0(off)

Low alarm

J

1(on) or 0(off)

Battery alarm

K

1(on) or 0(off)

Power off alarm

When you open the CSV file it will display readings taken every minute for the date chosen. 

The data will show refrigerator data performance in addition to the temperature including 

evidence of mains failure and extended door openings.

The columns read from left to right. The third column, column C, is for a second probe. If no 

probe is installed this  column will read at absolute zero (-273.1) which is normal and nothing 

to worry about. 

The codes for the other columns are as follows:

Содержание IntelliCold

Страница 1: ...www labcold com IntelliCold Pharmacy Refrigerator Operating Instructions...

Страница 2: ...irective and remain the user s responsibility to dispose of This refrigerator contains a flammable gas It MUST NOT be located in a room or area with a volume less than 10m3 This refrigerator contains...

Страница 3: ...14 Digital Door Lock 15 Caring for your fridge 18 Loading 18 Cleaning 19 Defrosting 19 Trouble Shooting 20 Useful Information 22 IMPORTANT These operating instructions contain information essential to...

Страница 4: ...during manufacture and dry thoroughly Ensure your fridge is level by adjusting the feet at the bottom of your fridge For models RLDF G0501A 1010A and 1510A insert shelves by placing clips in the rack...

Страница 5: ...ber on the display will descend accordingly ALTHOUGH THE REFRIGERATOR WILL REACH 5oC AFTER AN HOUR APPROX YOU SHOULD LEAVETHE FRIDGE RUNNING FOR 24 HOURS BEFORE FILLINGWITHTEMPERATURE SENSITIVEVACCINE...

Страница 6: ...asons for this such as extremely cold items being placed in the fridge The cause of the drop in temperature should be investigated immediately and the refrigerator should be monitored to check that it...

Страница 7: ...uter and or print hard copy The information stored in the controller will cover the last 13 months and is not deleted from the fridges memory when downloaded It is recommended you perform a download o...

Страница 8: ...n you have the correct day The display will then change to allow you to set the time Firstly you need to set the hour by using the buttons on the right of the display Press the tick button when you ha...

Страница 9: ...s all the temperature data copied to it since the timer was first set on the refrigerator or the last time the data was downloaded from the controller The Labcold SD card can now be inserted into a SD...

Страница 10: ...ntinually displays the first two digits of the model number ie 02 with no decimal point or the display does not change to an alarm state or show a temperature value after a few minutes there is a syst...

Страница 11: ...around LEDS Red Blue Place the battery terminals on the correct contact points PP3 Battery NOT included Opening the door Once the refrigerator has power you can now open the door using the factory set...

Страница 12: ...ion is paid to the loading of the fridge and it is regularly cleaned and defrosted If you are storing vaccines in your IntelliCold pharmacy refrigerator this should be recorded in accordance with reco...

Страница 13: ...vaccines If you press the set button on the front it will show the temperature the fridge is set to Should you wish to change the set point please contact the Labcold service team at service labcold...

Страница 14: ...idge is questioned by an official body Labcold are a UKAS calibration laboratory No 8898 and are therefore capable of providing calibration to ISO 17025 standards Why not contact our service team at s...

Страница 15: ...www labcold com...

Отзывы: