Only  use  original  filters  and  accessories  recommended 
by the manufacturer.  

  

 
 

8.

 

Trouble shooting 

 
 

Problem 

Possible cause 

Corrective action 

 
 
 
 
 
 
 
Suction capacity is low or 
the device does not aspire 

-Low motor speed selected 
 
-Valve could be blocked 
 
 
 
 
-Filter is dirty or wet  
 
-Battery defective or too low 
 
-Pipette holder is damaged 
 
-Sterile nose cone not tightly 
screwed 
 

-Increase motor speed  
 
-Clean the valve by rinsing. Dry it 
before installing. If the valve has 
been displaced, try to set it in the 
right position. 
 
-Replace filter 
 
-Charge batteries or replace them 
 
-Replace it 
 
-Tighten  the nose cone 

 
Pipette drips 

-Pipette not securely positioned 
in the silicon pipette holder 
 

-Attach again firmly 
 

 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 

Содержание EASY 60+

Страница 1: ...Electronic pipette filling device EASY 60 Please read the User Manual carefully before use and follow all operating and safety instructions user manual english espa ol...

Страница 2: ...he customer service representative with the following information Serial number on the back side Description of the problem Your contact information Warranty This instrument is guaranteed to be free f...

Страница 3: ...e possible risks However it is also important that you are alert for unexpected incidents 2 Intended use This Pipette Controller is an engine powered pipetting aid designed for cordless work with glas...

Страница 4: ...that isn t damaged and has no dents or sharp edges in the upper part Check that the upper gripping part is dry Hold the pipette carefully and insert it securely in to the silicon holder inside the co...

Страница 5: ...is low and the device needs to be charged The original charger supplied has protection against the overcharging short circuit and overheating Ensure that the voltage matches with the charger specific...

Страница 6: ...he batteries compartment cover pressing by embossed arrow and simultaneously sliding it back Remove the batteries Insert new batteries and reverse the above procedure Put the batteries compartment and...

Страница 7: ...ocked Filter is dirty or wet Battery defective or too low Pipette holder is damaged Sterile nose cone not tightly screwed Increase motor speed Clean the valve by rinsing Dry it before installing If th...

Страница 8: ...rsonal de Atenci n al Cliente la siguiente informaci n N mero de serie del equipo en el panel trasero Descripci n del problema detectado Sus datos de contacto Garant a Este equipo est garantizado cont...

Страница 9: ...mportante que est alerta ante incidentes inesperados 2 Uso Este controlador de pipeta es un auxiliar de pipeteo motorizado dise ado para trabajo inal mbrico con pipetas de vidrio o pl stico en el rang...

Страница 10: ...afilados en la parte superior Compruebe que la parte superior del agarre est seca Sostenga la pipeta con cuidado e ins rtela de manera segura en el soporte de silicona dentro del cono Una vez que la...

Страница 11: ...ignifica que el nivel de la bater a es bajo y que el dispositivo debe cargarse El cargador original suministrado tiene protecci n contra sobrecarga cortocircuito y sobrecalentamiento Aseg rese de que...

Страница 12: ...n relieve y al mismo tiempo desliz ndola hacia atr s Retire las pilas Inserte bater as nuevas y revierta el procedimiento anterior Coloque el compartimiento de las bater as y vuelva a colocar la tapa...

Страница 13: ...o El cono de sujeci n no est bien fijado Seleccione una velocidad de trabajo superior Limpie la v lvula mediante enjuague Aseg rese que est seca antes de volverla a instalar Si la v lvula se ha despla...

Страница 14: ...e en vertederos De conformidad con la Directiva 2002 96 CE los usuarios de la Uni n Europea de aparatos el ctricos y electr nicos tienen la oportunidad de retornar el instrumento para su eliminaci n a...

Страница 15: ...www labbox com...

Отзывы: