adjust settings. Wrong operation may cause abnormal working of the device,
such as bad temperature uniformity and abnormal alarm.
Parameter Settings
When power on the medical refrigerator, the display screen displays internal
temperature value. To change setting temperature, for example, the original
set temperature is 2
℃
which now has to be adjusted to 5
℃
, please operate
according to the following steps:
Serial
Keys Operation
Display
No.
1
Displays the inside temperature
2
Press the key 8 for 3 seconds
Displays the initial set data is
2
℃
and twinkles
3
Press the key 9
Figure increases
4
Press the key 10
Figure decreases
5
Figure is 5
℃
6
Press the key 8 for 3 seconds
Saves the user settings and
displays the inside temperature
Notes:
If other parameter need to be changed, please contact after-sale or
local distributor of us for setting.
Temperature Query Settings
When normal working, press key 9 check high and low temperature record.
Such as current time is 9:20on the afternoon of Mar 15th, 2017 record the
high temperature of 6.7
℃
, at 15;23 record the low temperature of 3.9
℃
, and
the cycle. No operation is larger than that of 5S showed normal temperature.
The query is shown below:
NO.
key operation
Display
1
Display cabinet temperature
2
Press key 9
Display highest temperature6.7
℃
3
Press key 9 again
Display HT date HD20
4
Press key 9 again
Display HT hour HH09
5
Press key 9 again
Display HT minute HE30
6
Press key 9 again
Display lowest temperature3.9
7
Press key 9 again
Display LT date LD20
8
Press key 9 again
Display LT hour LH14
9
Press key 9 again
Display LT minute LE25
10
Press key 9 again
Display highest temperature6.7
℃
5