42
MAINTENANCE:
(Con’t)
STAINLESS CLEANING METHODS: (Con’t)
•
Bathroom tub and tile cleaners.
•
A mixture of 20% nitric acid and 1.5% hydrofluoric acid (or hyrochloric
acid). Swab solution on surface allowing it to remain until rust is loosened.
Immediately follow with a clean water rise. (This method should ONLY be
used if SEVERE rust and scale stains are present.)
•
Oxalic acid 2% to 5% in warm water. Swab solution on surface allowing it
to remain until rust is loosened. Immediately follow with a clean water
rise. (This method should ONLY be used if SEVERE rust and scale stains
are present.)
Regardless of the approach utilized, ALWAYS follow the manufacturer’s
directions and allow the chemicals to do the cleaning with MINIMAL scrubbing.
Always follow cleanings with a clean water rinse. Air dry.
MATERIALS EFFECTIVE IN DISINFECTING STAINLESS:
•
Glutaraldehyde
•
Alcohol
BE
ADVISED:
THIS INFORMATION IS INTENDED AS
GUIDELINES ONLY
AND LAB-LINE INSTRUMENTS, INC. MAKES NO CLAIM AS TO THE
SUITABILITY TO ANY PARTICULAR SITUATION. CONSULT YOUR STAFF
CHEMIST TO DETERMINE WHAT WOULD BE BEST FOR YOUR STAINLESS
STEEL PRODUCT AND LABORATORY.
CLEANING COPPER SURFACES: FLO-THRU MODELS, 316/317 SERIES:
Copper surfaces can be cleaned using commercially available detergent
preparations specifically indicated for copper use. Copper surfaces will not
tarnish when washed with detergent solutions that contain small amounts of the
chemical compound 1-, 2- or 3-benzotriazole. It is important, however, before
using the detergents to determine any possible effect the presence of this
compound and any other ingredients might have on the incubation work being
conducted. Contact the manufacturers of the commercially available detergents
and obtain all necessary information on the chemical and physical properties of
the compositions.
TEMPERATURE READOUT CALIBRATION: ALL MODELS:
Place a scientifically accurate thermometer in the chamber, on a center
shelf and allow the incubator to stabilize at some set temperature. If the digital
temperature readout does not agree with the thermometer, take the trimmer pot
tool, insert in calibration pot on control panel and adjust until readout matches
the measured temperature.
UNIT REPAIRS:
In the event that your unit becomes inoperative, consult with your local
dealer for assistance.