Operation

Heated Operation

For faster performance, preheat the unit by turning the “HEAT POWER ON” switch to the right and 
setting the temperature controller to the desired operating temperature. In some conditions, it may 
be necessary to set the operating temperature approximately 15-20 degrees above the desired 

sample temperature (maximum 160⁰F). Dry block or water bath heaters may be purchased 

separately to preheat samples.

The temperature probe is located underneath the lid, above the centrifuge bowl. The temperature 
display indicates the temperature inside the bowl, NOT the samples. To determine sample 
temperature, a thermometer must be placed inside each sample tube once a test is complete. 
Thermometers may be purchased separately. 

Motor Operation

• Press the “PUSH TO RUN” button. Verify the motor RPMs in the display, which shows actual RPMs

 (not theoretical).

• Turn the “SPEED ADJUST” control knob to the desired speed (see Appendix A for various RCF

values). To prolong centrifuge life, use the lowest feasible speed setting. The maximum allowable 
speed is 2000 rpm. DO NOT attempt to run the motor with the door open, as the unit is designed to 
shut off when the door is opened.

• When the RPM INDICATOR reads “0”, wait a few seconds to open the door. DO NOT force the door 

open. A locking mechanism automatically engages the when the motor is turned on and stays 
engaged until the motor stops spinning. 

• Verify that th

e trunnion arm has come to a complete stop. Remove samples and close the door. 

Keep the door closed at all times, as this minimizes heat loss.

• Contact the L-K 

Industries Tech Helpline for instructions for manually bypassing the locking 

mechanism. The manual bypass should ONLY be done to verify centrifuge speed.

Timer Operation

The Countdown Timer Display indicates the amount of remaining time on a test. If the countdown 

time must be changed, the controller must be reprogrammed. The controller will prompt for a code 

to change the program. Enter code “055”. See Appendix C for the Timer Quick Programming Guide. 
Otherwise, contact the L-K Industries Tech Helpline for timer programming instructions

NOTE: 

On initial startup or after any programming changes, reset the unit before beginning operation.

Содержание BENCHMARK S

Страница 1: ...MANUFACTURING QUALITY and customer service in the PETROLEUM TESTING industry BENCHMARK S MANUAL Heated Laboratory Centrifuge Demand a higher standard L K INDUSTRIES Petroleum Testing Equipment Supplie...

Страница 2: ...ately following discovery of defect s Defective products under warranty shall only be returned after contacting and receiving permission from L K Industries The warranty does not extend to L K Industr...

Страница 3: ...Installation Electrical Connection 8 9 9 9 Loading and Powering On Heated Operation Motor Operation Timer Operation MAINTENANCE 10 10 10 Replacement Parts Repair Services Housekeeping Replacement of F...

Страница 4: ...the limits of this unit will help prevent accidents REMOVE all foreign objects from the centrifuge bowl Keep the sliding door CLOSED while samples are spinning NEVER try to slow the trunnion arm down...

Страница 5: ...rifuge should be housed in a controlled environment It is recommended that the centrifuge be secured to the benchtop table stand etc using the hardware supplied Allow for optimum working clearance aro...

Страница 6: ...nit is not hard wired a power cord with the appropriate connections shall ONLY be purchased through the L K Industries warehouse In geographical regions where power load is prone to uctuation power co...

Страница 7: ...N DISCONNECT switch to the right The green indicator light should signify that power is on Visually verify that the EMERGENCY MOTOR OFF SWITCH is not engaged Open the sliding door and place the lled s...

Страница 8: ...the lowest feasible speed setting The maximum allowable speed is 2000 rpm DO NOT attempt to run the motor with the door open as the unit is designed to shut o when the door is opened When the RPM INDI...

Страница 9: ...e trunnion shields Before each test examine moving parts for wear and stress Replace parts as necessary If a tube has broken inside a shield the cushion MUST be replaced Glass particles can become emb...

Страница 10: ...the OFF position If the motor does not run ensure that the door is properly closed The proximity switch uses a sensor to determine whether the door is closed and closes a circuit allowing power to fe...

Страница 11: ...Door and Locking Mechanism Assembly Front Panel and Lid Assembly Appendix C Programming Overviews The following programming diagrams may be found at www lk ind com downloads RPM Quick Programming Time...

Страница 12: ...ger Midland Scienti c Parkes Scienti c www amazon com www coleparmer com www dcscienti c com www grainger com www midlandsci com www parkesscienti c com Catalog 2016 L K Industries All Rights Reserved...

Отзывы: