Data acquisition start
synchronization
Implemented
in version
E20-10
Notes
"external - slave on START
signal edge":
- without additional
conditions
- with additional start
conditions.
A, V
V
Start on START signal
Stopping possible conditions -
"external – slave on START
signal fall":
- without additional
conditions
- with additional start
conditions.
A, V
V
Start on START signal
Possible stopping conditions see in
"on level" data acquisition
synchronization modes,
V
On signal analog level in selected channel
Table 5-3. Capability to set additional permit conditions for ADC data recording in
buffer memory E20-10:
Additional conditions
Implemented in
version E20-10
Notes
Blockage of given
N
number of frames recording
from the acquisition start
B
0 ≤
N
≤ 16777214,
where, N-integral
On bottomup transition through the given signal level
in selected channel and blockage of
N
number of
frames recording from acquisition start
B
0 ≤
N
≤ 16777214,
where,
N
-integral
On top-down transition through the given signal level
in selected channel and blockage of
N
number of
frames recording from acquisition start
B
0 ≤
N
≤ 16777214,
where,
N
-integral
Table 5-4. "On level" synchronization modes
"on level" acquisition modes
(acquisition start and stop is performed on signal
level)
Impleme
nted in
revision
E20-10
Notes
"on signal level above threshold of signal in 1 ADC channel"
B
Stopping will occur at
frame end edge
"on signal level above threshold of signal in 2 ADC channel"
B
"on signal level above threshold of signal in 3 ADC channel"
B
"on signal level above threshold of signal in 4 ADC channel"
B
"on signal level below threshold of signal in 1 ADC channel"
B
"on signal level below threshold of signal in 2 ADC channel"
B
"on signal level below threshold of signal in 3 ADC channel"
B
"on signal level below threshold of signal in 4 ADC channel"
B