6.1.2.4. Differential winding connection
with midpoint
6.1.2.5. Differential winding connection
with midpoint through the divider
Locate R1 close to E-502 input.
Only for E-502 voltage subbands:
±2 V, ±1
V, ±0.5 V, ±0.2 V
Locate R1 close to E-502 input.
Only for E-502 voltage subbands:
±2 V, ±1
V, ±0.5 V, ±0.2 V
6.1.3. Connection to the ADC input for the case where the common wire of the
signal sources has a offset potential
Ucm of max. ± 1 V
relative to the AGND
circuit.
6.1.3.1. Connection up to 32 channels.
Mode "with common ground"
6.1.3.2. Connection up to 16 channels.
"Differential" mode
See limitations in section 4.6, p. 412!
See limitations in section 4.6, p. 412!
6.1.4. Measurement of the voltage drop on the circuit section in the differential
mode (up to 16- channels)
This connection allows you to measure the voltage drop
across resistor R2.
For a single-channel mode
, R1, R2, R3
can also be impedances of a capacitive or inductive nature,
but with the condition that the circuits X and Y used are not
broken by the direct current of the circuit.
In multichannel
mode
, the equivalent impedance referred to the ADC input
must be active. It is necessary that R2 should not be more
than 50 ohms if the switching frequency is maximal.
See
limitations in section 4.6, p.412!
R1
E-502
AGND
X1 (X2 -X16)
Y1 (Y2 -Y16)
L
R1
E-502
AGND
X1 (X2 -X16)
Y1 (Y2 -Y16)
R2
R3
L
E-502
X2 (X3-X16, Y1-Y16)
AGND
GND32
U
вх
1
U
см
X1
U
вх
2 (3-16)
E-502
AGND
U
вх
1
U
см
U
вх
2 (3-16)
Y1 (Y2 -Y16)
X1
X2 (X3 -X16)
R2
E-502
AGND
X1 (X2 -X16)
Y1 (Y2 -Y16)
R3
R1