When plugging your finger into the Oximeter, your nail surface must be upward.

Declaration: Please use the medical alcohol to clean the rubber before each test and clean the

tested finger with alcohol before and after the test. (The rubber inside of the Oximeter adopts

medical rubber, which has no toxin, no harm, and brings no side effect such as allergy to the

our skin ).

Brief Description of Front Panel

Key function

start machine in power off status; change display direction in working status.

Detailed descriptions of product functions:

1.

Display Type

LCD display or OLED display

2.

SpO2

Measurement range

70%

99%

Accuracy

:±

2% on the stage of 70%

99%

Unspecified(

70%) for SPO2

Resolution

:±

1%

3.

PR

Measurement range

30BPM

240BPM

Accuracy

:±

1BPM or

±

1%

the larger one

4.

Power

two AAA 1.5V alkaline batteries

5.

Power consumption

below 30mA

6.

Automatic power-off

the product shuts off by itself when no finger is in the product in 8

seconds

7.

Dimension

62mm

×

32mm

×

33mm

8.

Operation Environment

Operation Temperature

5

℃~

40

Storage Temperature

-10

℃~

40

Ambient Humidity

15%

80% on operation

10%

80% in storage

Air Pressure

70kPa

106kPa

9.

Declaration

EMC of this product comply with IEC60601-1-2 standard.

10. Measurement Performance in Low Perfusion Condition

required test equipment

(BIO-TEK INDEX Pulse Oximeter tester) could measure the available pulse wave with the

amplitude of 6% of the simulation pulse wave amplitude.

11. 11. Interference Resistance Capacity against Ambient Light:

Device works normally when

BIO-TEK INDEX Pulse Oximeter tester exerts interfering signal test.

Classification

1.

Management Class for Medical Devices:

equipment

2.

Anti-electric Shock Type: Internally powered equipment

3.

Anti-electric Shock Degree: Type BF equipment

Maintenance and Preservation

1.

Replace the batteries timely when low voltage lamp is on.

2.

Clean the surface of fingertip oximeter before it is used to diagnose patients.

3.

Remove the batteries inside if you will not operate the Oximeter for a long time.

4.

It would be better to preserve the product in -10~40

(14-104

) and humidity is

10%-80%.

5.

It is recommended that the product should be kept dry anytime. A wet ambience might

affect its lifetime and even damage the product.

6.

Please follow the law of the local government to deal with used batteries.

Product Accessories

1.

One hang lace

2.

Two batteries

3.

One user manual

Guidance and manufacture’s declaration-electromagnetic radiation-for other
EQUIPMENTS and SYSTEMS

The Pulse Oximeter is designed to be used in specified electromagnetic environment .

Users of the Pulse Oximeter must use it in the following environments.

Radiation Test

Compliance

Electromagnetic environment-guidance

RF interference

CISPR 11

Group 1

RF signal of Pulse Oximeter is simply created by
its

internal

function.

Therefore,

its

RF

interference is very low and is not likely to cause
any interference to nearby electronic equipment.

RF interference

CISPR 11

Class B

The Pulse Oximeter applies to all establishments,
including domestic establishments and those
directly connected to the public low-voltage power
supply network that supplies buildings used for
domestic purposes.

Possible Problems and Resolutions

Problem

Possible reason

Solution

SpO2 or PR

can not be

shown
normally

1. Finger is not plugged correctly
2.Patient’s Oxyhemoglobin value

is too low to be measured

1. Retry by plugging the finger

2. Try more times. If you can

make sure there is no problem in

the product, please go to hospital

timely for exact diagnosis

SpO2 or PR

is

shown

unsteady

1. The finger might not be plugged

deep enough

2. Finger is trembling or the

patient is on movement status

1. Retry by plugging the finger
2. Please remain at rest

The

Oximeter

can not be
turned on

1. Inadequate power or power off

2. Batteries might be installed

incorrectly

3.

The

Oximeter

might

be

damaged

1. Please replace the batteries

2. Please reinstall the batteries

3. Please contact with local

customer service centre

Indication

lamps

are

suddenly off

1. The product automatically shuts

off when no signal is detected in 8
seconds

2. Inadequate power

1. Normal
2. Replace the batteries

Symbols and Definitions

Symbol

Definition
The equipment type is BF

Refer to user manual before application

SpO2

Hemoglobin Saturation

PR

Heart Rate(BPM)

Low power indication

SN

Serial No.

LCD display:

LCD display modes

OLCD display:

OLCD display modes

1.

2.

3.

4.

start/mode key

start/mode key