KYTO KYTO80A Скачать руководство пользователя страница 1

User Manual

Ver.1.0

Fingertip Pulse Oximeter

This Fingertip Pulse Oximeter is a kind of innovated medical device with non-invasive and

continuous features for artery SPO2 and PR detection. Being portable, it is able to measure

SPO2 and PR values quickly and precisely.

General Description

Haemoglobin Saturation is the percentage between the capacity of Oxyhemoglobin (HbO2)

that compounded with oxygen and that of all combinativable haemoglobin (Hb) obin (HbO2)

in blood. In other words, it is the saturation of Oxyhemoglobin in blood. It is a very important

physiological parameter for Respiratory and Circulation Systems. Many respiratory diseases

could reduce haemoglobin saturation in human blood. Moreover, factors such as Automatic

Organic Regulation Malfunction caused by anaesthesia, trauma resulted from major operation

and some medical examination can also cause problems in oxygen supply, which might reduce

human haemoglobin saturation. As a result, such symptoms as megrim, vomiting and asthenia

might appear to patients. Hence, it is very important to know hemoglobin saturation of patient

timely in clinical medical aspects.

The fingertip pulse oximeter features in small volume, low power consumption, convenient

operation and

portability. It is only necessary for patient to put one finger into fingertip

photoelectric sensor for diagnosis, and the display screen will directly show measured value of

hemoglobin saturation. It has been proved in clinical experiments that it possesses rather high

precision and repeatability.

Measurement principle

The principle of the oximeter is as follows: An experience formula of data process is

established by exerting Lambert Beer Law according to Spectrum Absorption Characteristics

of Reductive hemoglobin(R Hb) and oxyhemoglobin (O2 Hb) in glow and near-infrared zones.

Operation principle of the instrument is to combine Photoelectric Oxyhemoglobin Inspection

Technology with Capacity Pulse Scanning and Recording Technology, so that two lights with

different wavelength (660nm glow and 940nm near infrared light) can be focused onto human

nail through perspective clamp finger-type sensor. Then measured signal can be obtained by a

photosensitive element, information acquired through which will be shown on two groups of

LEDs through process in electronic circuits and microprocessor.

Diagram of Operation Principle

1. Infrared-ray receiving tube

2. Infrared-ray transmitting tube

Precautions for use

1.

Do not use the pulse oximeter together with MRI or CT equipment.

2.

Explosion hazard: Do not use the pulse oximeter in an explosive atmosphere.

3.

The pulse oximeter is intended only as an adjunct in patient assessment. Doctors should

make diagnosis in conjunction with clinical manifestation and symptoms.

4.

Check the pulse oximeter sensor application site frequently to make sure that the

circulation and skin integrality of patient are under good condition.

5.

Do not stretch the adhesive tape while applying the pulse oximeter sensor. This may

cause inaccurate reading or skin blisters.

6.

Please read the manual carefully before your operation.

7.

The pulse oximeter has no SpO2 alarms, it is not for continuous monitoring.

8.

Prolonged use or the patient’s condition may require changing the sensor site periodically.

Change sensor site and check skin integrality, circulatory status, and correct alignment at

least every 4 hours.

9.

Inaccurate measurements may be caused by autoclaving, ethylene oxide sterilizing, or

immersing the sensors in liquid.

10. Significant levels of dysfunctional hemoglobins (such as carboxyl- hemoglobin or

methemoglobin) may cause inaccurate reading.

11. Intravascular dyes such as indocyanine green or methylene blue may cause inaccurate

reading.

12. SpO2 measurements may be adversely affected in the presence of high ambient light.

Please shield the sensor area (with a surgical towel or direct sunlight, for example) if it is

necessary.

13. Unexpected action may cause inaccurate reading.

14. Medical signal with high frequency or interference caused by defibrillator may lead to

inaccurate reading.

15. Venous pulsations may cause inaccurate reading.

16. It may cause inaccurate reading when the positions of sensor and blood pressure cuff are

on the same arterial catheter or intravascular line.

17. Hypotension, severe vasoconstriction, severe anemia, or hypothermia may cause

inaccurate reading.

18. It may cause inaccurate reading by giving use of cardiotonic to patient after his cardiac

arrest or when he is in quiver.

19. Bright nail or painted nail may cause inaccurate SpO2 reading.

Follow local ordinances and recycling instructions regarding to disposal or recycling of the

device and device components, including batteries.

Features

LCD display

LCD display, easy to operate

Low-power consumption, continuously work for 40 hours

Low voltage indicator

Automatically power off in 8 seconds when there is no signal

Small in volume, light in weight, and convenient to carry

OLED display

Two color OLED display, more display modes

Low-power consumption, continuously work for 40 hours

Low voltage indicator

Automatically power off in 8 seconds when there is no signal

Small in volume, light in weight, and convenient to carry

Product Operation Scope

The fingertip Oximeter can be used to measure human haemoglobin saturation and heart rate

through finger. The product applies to using in family, hospital (including clinical use in

internist/surgery, anaesthesia, paediatrics, intensive care, etc.), oxygen club, social medical

organizations, physical care in sports (It can be used before or after sports. Operation in sport

procedure is not recommended). It is also applicable to enthusiasts on mountaineering,

patients (convalescents at home or those need first aid treatment), elders over 60, those work

more than 12 hours, sporters and those work in the hermetic circumstance, etc. The product is

not suitable to monitor patient continuously.

Operation Instructions

1.

Install two AAA batteries into battery cassette before covering its cover.

2.

Plug one finger into rubber hole of the Oximeter (it is best to plug the finger thoroughly)

before releasing the clamp with the nail upwards.

3.

Press button on the front panel;

4.

Don’t tremble your finger when the Oximeter is working. Your body is not recommended

on moving status.

5.

Press the button on the front panel, if we want change display direction;

6.

Read relevant datum from display screen.

7.

If there is no signal input, oximeter can shut off automatically

8.

Please replace new batteries when OLED indicates the batteries are in low power.

Отзывы: