
User Manual
Ver.1.0
Fingertip Pulse Oximeter
This Fingertip Pulse Oximeter is a kind of innovated medical device with non-invasive and
continuous features for artery SPO2 and PR detection. Being portable, it is able to measure
SPO2 and PR values quickly and precisely.
General Description
Haemoglobin Saturation is the percentage between the capacity of Oxyhemoglobin (HbO2)
that compounded with oxygen and that of all combinativable haemoglobin (Hb) obin (HbO2)
in blood. In other words, it is the saturation of Oxyhemoglobin in blood. It is a very important
physiological parameter for Respiratory and Circulation Systems. Many respiratory diseases
could reduce haemoglobin saturation in human blood. Moreover, factors such as Automatic
Organic Regulation Malfunction caused by anaesthesia, trauma resulted from major operation
and some medical examination can also cause problems in oxygen supply, which might reduce
human haemoglobin saturation. As a result, such symptoms as megrim, vomiting and asthenia
might appear to patients. Hence, it is very important to know hemoglobin saturation of patient
timely in clinical medical aspects.
The fingertip pulse oximeter features in small volume, low power consumption, convenient
operation and
portability. It is only necessary for patient to put one finger into fingertip
photoelectric sensor for diagnosis, and the display screen will directly show measured value of
hemoglobin saturation. It has been proved in clinical experiments that it possesses rather high
precision and repeatability.
Measurement principle
The principle of the oximeter is as follows: An experience formula of data process is
established by exerting Lambert Beer Law according to Spectrum Absorption Characteristics
of Reductive hemoglobin(R Hb) and oxyhemoglobin (O2 Hb) in glow and near-infrared zones.
Operation principle of the instrument is to combine Photoelectric Oxyhemoglobin Inspection
Technology with Capacity Pulse Scanning and Recording Technology, so that two lights with
different wavelength (660nm glow and 940nm near infrared light) can be focused onto human
nail through perspective clamp finger-type sensor. Then measured signal can be obtained by a
photosensitive element, information acquired through which will be shown on two groups of
LEDs through process in electronic circuits and microprocessor.
Diagram of Operation Principle
1. Infrared-ray receiving tube
2. Infrared-ray transmitting tube
Precautions for use
1.
Do not use the pulse oximeter together with MRI or CT equipment.
2.
Explosion hazard: Do not use the pulse oximeter in an explosive atmosphere.
3.
The pulse oximeter is intended only as an adjunct in patient assessment. Doctors should
make diagnosis in conjunction with clinical manifestation and symptoms.
4.
Check the pulse oximeter sensor application site frequently to make sure that the
circulation and skin integrality of patient are under good condition.
5.
Do not stretch the adhesive tape while applying the pulse oximeter sensor. This may
cause inaccurate reading or skin blisters.
6.
Please read the manual carefully before your operation.
7.
The pulse oximeter has no SpO2 alarms, it is not for continuous monitoring.
8.
Prolonged use or the patient’s condition may require changing the sensor site periodically.
Change sensor site and check skin integrality, circulatory status, and correct alignment at
least every 4 hours.
9.
Inaccurate measurements may be caused by autoclaving, ethylene oxide sterilizing, or
immersing the sensors in liquid.
10. Significant levels of dysfunctional hemoglobins (such as carboxyl- hemoglobin or
methemoglobin) may cause inaccurate reading.
11. Intravascular dyes such as indocyanine green or methylene blue may cause inaccurate
reading.
12. SpO2 measurements may be adversely affected in the presence of high ambient light.
Please shield the sensor area (with a surgical towel or direct sunlight, for example) if it is
necessary.
13. Unexpected action may cause inaccurate reading.
14. Medical signal with high frequency or interference caused by defibrillator may lead to
inaccurate reading.
15. Venous pulsations may cause inaccurate reading.
16. It may cause inaccurate reading when the positions of sensor and blood pressure cuff are
on the same arterial catheter or intravascular line.
17. Hypotension, severe vasoconstriction, severe anemia, or hypothermia may cause
inaccurate reading.
18. It may cause inaccurate reading by giving use of cardiotonic to patient after his cardiac
arrest or when he is in quiver.
19. Bright nail or painted nail may cause inaccurate SpO2 reading.
Follow local ordinances and recycling instructions regarding to disposal or recycling of the
device and device components, including batteries.
Features
LCD display
LCD display, easy to operate
Low-power consumption, continuously work for 40 hours
Low voltage indicator
Automatically power off in 8 seconds when there is no signal
Small in volume, light in weight, and convenient to carry
OLED display
Two color OLED display, more display modes
Low-power consumption, continuously work for 40 hours
Low voltage indicator
Automatically power off in 8 seconds when there is no signal
Small in volume, light in weight, and convenient to carry
Product Operation Scope
The fingertip Oximeter can be used to measure human haemoglobin saturation and heart rate
through finger. The product applies to using in family, hospital (including clinical use in
internist/surgery, anaesthesia, paediatrics, intensive care, etc.), oxygen club, social medical
organizations, physical care in sports (It can be used before or after sports. Operation in sport
procedure is not recommended). It is also applicable to enthusiasts on mountaineering,
patients (convalescents at home or those need first aid treatment), elders over 60, those work
more than 12 hours, sporters and those work in the hermetic circumstance, etc. The product is
not suitable to monitor patient continuously.
Operation Instructions
1.
Install two AAA batteries into battery cassette before covering its cover.
2.
Plug one finger into rubber hole of the Oximeter (it is best to plug the finger thoroughly)
before releasing the clamp with the nail upwards.
3.
Press button on the front panel;
4.
Don’t tremble your finger when the Oximeter is working. Your body is not recommended
on moving status.
5.
Press the button on the front panel, if we want change display direction;
6.
Read relevant datum from display screen.
7.
If there is no signal input, oximeter can shut off automatically
;
8.
Please replace new batteries when OLED indicates the batteries are in low power.