5-3 INSULATION CHECKING
Push
INPUT
key to display the mode selection window.
1)
Blinking
Select
<INSUL>
with the cursor key.
2)
Blinking
Push the
ENTER
key for setting.
3)
<
INSUL
> and insulation resistance value appear on the
4)
monitor window.
5) Connect the input cable to the INPUT terminal.
6) Connect the alligator clip and white clip of the input cable
to the target place.
[Connection Method]
Red (+BV)
White (-SIG)
Black (-BV)
Green (+SIG)
Crocodile clip (Shield)
23
Insulation resistance value should be within 2 to 100 Mohm.
MEMO
If the target insulation resistance value is lower than 2 Mohm, <
LOW
> appears, higher than 100
Mohm, <
HIGH
> appears.