— 16 —
6.2 Class I Test (Select the Leakage current test)
Selecting Class I test while Select Switch is being pressed down initiates
Leakage current test instead of Insulation resistance test.
In case of selecting Leakage current test, metal parts other than the heating
or movable parts must be clipped with Test Lead M-7129A since the DUT
activates. Pressing the START/STOP switch during Leakage current test
stops the test immediately. To restart the test, press the START/STOP
switch again 2 sec or later than the stop of the test. Then (1) Protective
conductor resistance will be restarted.
Fig 12 indicates how to connect the devices.
Class I Test (Select the Leakage current test) Flowchart
Start
(
1
). Protective conductor resistance.
RPE <
=
1
Ω
?
(
2
). Power on Leakage current test.
LEA <
=
5
mA?
Light up in red
no
→
Con
→
value will
be repeated on the LCD.
Light up in red
Stop the test and
no
→
LEA
→
value will
be repeated on the LCD.
Value of (
1
). and (
2
). will be
alternately displayed on the LCD.
No
No
Yes
Yes
PASS
Light up in green
Light up in green
Brief indications S- and on will be displayed
in turns on the LCD for
5
sec and prompt the
user to set the DUT switch on.
Then a mark indicating working DUT function
will be displayed for
10
sec.
5
sec
10
sec