3. Specification
● Measurement range and accuracy (23ºC±5Cº, relative humidity
75% or less)
Voltage/ Earth voltage measurement
Measurement range
Display range
Accuracy
AC5.0 to 300.0V
(45 to 65Hz)
(425V peak or less)
0.0 to 314.9V
±1%rdg±4dgt
DC±5.0 to ±300.0V
0.0 to ±314.9V
±1%rdg±8dgt
*AC measurement method : rms detection
For the waveforms other than sinewave
of CF<2.5, the specified accuracy
±1%f.s. should be applied.
* AC/DC auto-detection at 5V or higher
* DC sign appears at 5V or higher.
(H+S(C+P) terminal: positive side, E terminal : negative side)
* The LCD reads 0.0V for 0.9V or lower voltages.
Earth resistance measurement
Range
(auto-ranging)
Measurement
range
Display range
Accuracy
200Ω Range
5.0 to 2000Ω
0.0 to 209.9Ω
±3%rdg±5dgt
2000Ω Range
160 to 2099Ω
Measurement method: Constant current inverter
approx. 1.4mA (200Ω Range)/825Hz
approx. 0.7mA (2000Ω Range)/825Hz
Open voltage: Approx. 13V