MODEL3323A
* Heavy-line circular arc on the scale indicates the primary effective
measuring ranges. (guaranteed accuracy range)
<Operating error>
Operating error (B) is an error obtained under the nominal
operating conditions, and calculated with the intrinsic error (A),
which is an error of the instrument used, and the error (En) due to
variations. According to IEC61557, the maximum operating error
should be /-30%.
B =︱A︱+ 1.15 ×
√
( E
12
+ E
22
+ E
32
)
A : Intrinsic error (%)
B : Operating error (%)
E
1
: Variation due to changing the temperature (%)
E
2
: Variation due to changing the Battery voltage (%)
E
3
: Variation due to changing the position (%)
Nominal operating conditions
Ambient temperature : 0
〜
40℃
Relative humidity
: 90% or less
External magnetic field : 400A/m or less
Position
: Horizontal
〜
±90°
Battery voltage
: within "BATTERY. GOOD" range
Nominal voltage
Max. effective scale value
Mid-scale value
Accuracy in primary
effective measuring ranges
Accuracy in primary
effective measuring ranges
Accuracy at 0 and ∞
Accuracy at no-load
voltage
Nominal current
Short-circuit current
25V
10MΩ
0.2MΩ
0.01〜5MΩ
within ±5% of indicated value
Measuring ranges other than above, 0 and ∞.
within ±10% of indicated value
within ±0.7% of scale length
0%〜+20% of nominal voltage
1mA 0%〜+20%
within 1.5mA
50V
10MΩ
0.2MΩ
0.01〜5MΩ
100V
20MΩ
0.5MΩ
0.02〜10MΩ