̶ 13 ̶
[How to attach]
Firmly insert and connect the adapter to the end of the cord (with
banana plugs at both ends).
#
DANGER
To avoid getting electrical shocks, ensure that test leads are
disconnected from the instrument when replacing the metal tip or
adapter for test leads.
6-3 Battery voltage check
(1) Please refer to 11. Battery replacement in this manual and
insert batteries in KEW3431.
(2) Set the function switch to any position other than OFF to power
on theinstrument.
(3) Check the color of LED battery status indicator.
Solid green : Battery voltage is enough.
Blinking green : Battery voltage is low.
Replace batteries with reference to 11. Battery replacement
to perform further measurements.
Solid red : Battery voltage is below the lower limit of the operating
voltage. In such a condition, accuracy of the measured result
isn t guaranteed. Replace the batteries with new ones as soon
as possible.
● Color of LED battery status indicator might change from green to
red depending on measured objects; for example, resistance of
the measured object is low.
● Use of AA alkaline battery is recommended. Use of other
batteries may cause improper indication of battery level.
Banana plugs, both ends
Fig. 6-2
Flat test bar
Alligator clip
Содержание KEW3431
Страница 1: ...Instruction Manual Compact Insulation resistance tester KEW3431 ...