— 5 —
― 4 ―
2.Features
MODEL3161A is a two-range insulation resistance tester for
testing various elevators and safety equipments.
¡
Designed to following safety standards:
IEC 61010-1, IEC 61010-031, IEC 61557 (500V/ 100MΩ only)
¡
Small and light weight.
¡
Auto-discharge function
When insulation resistance like a capacitive load is measured,
electric charges stored in capacitive circuits are automatically
discharged after measuring. Discharge can be checked with the
meter.
¡
Color-coded scales for easy and correct reading
¡
Power-on indicator on insulation resistance and battery check
ranges.
¡
Backlight function to facilitate working at dimly illuminated
location or at nighttime work.
¡
AC voltage measurement function
¡
Measures AC voltage on all ranges without pressing the Test
button
¡
Test leads with remote control switch
(voltage won't be outputted when the test lead is not connected.)
¡
Robust housing case
¡
Neck strap for both hand's operation
¡
User-changeable test prod
500V
100MΩ
2MΩ
0.1~
50MΩ
― 5 ―
15V
20MΩ
0.05MΩ
0.005~
2MΩ
within ±5% of indicated value
Measuring ranges other than above, 0 and ∞.
within ±10% of indicated value
within ±0.7% of scale length
0%~+20% of nominal voltage
¡
Applicable standards
IEC 61557 (except for 15V/ 20MΩ range)
IEC 61010-1 Measurement CAT.III 300V/ CAT.II 600V
Pollution degree2, Protection class II
Location for use: altitude 2000m or less
IEC 61010-031
IEC 60529
IP40
¡
Measuring range and accuracy
<Insulation resistance range>
3.Specification
Nominal voltage
Max. effective
scale value
Mid-scale value
Accuracy in
primary effective
measuring ranges
Accuracy in
secondary effective
measuring ranges
Accuracy at 0 and ∞
Accuracy at
no-load voltage
Nominal current
Short-circuit
current
Response time
※
Heavy-line circular arc on the scale indicates the primary
effective measuring ranges. (guaranteed accuracy range)
1mA 0%~+20%
Not specified
Approx. 2mA
Meter reading comes within accuracy within 3 sec.
(It may take time when measuring a capacitive load.
3.Specification
◦
Applicable standards
IEC 61557 (except for 15V/ 20MΩ range)
IEC 61010-1, IEC 61010-2-030
Measurement CAT III 300V/ CAT II 600V
Pollution degree2, Protection class II
Location for use: altitude 2000m or less
IEC 61010-031
IEC 60529
IP40
EN 61326-1 EMC
EN 50581 RoHS
◦
Measuring range and accuracy
<Insulation resistance range>
※
Heavy-line circular arc on the scale indicates the primary
effective measuring ranges. (guaranteed accuracy range)