15
Table 7.1(cont.).
1
2
3
4
12,14
AFC CAVITY
Toggle switch
–
discriminator down
RF cavity AFC unit switches on.
Multi-turned potentiometer sets RF
cavity frequencies initial position.
Down
13
INDICATION
Slide switch
–
switches "FRQ.ADJ"
down code or "TIME" code on
indicator.
Down
15
I
Button
–
ion pump current check.
16,18
CHECK
Switch
–
switches standard test points
to "CHECK" indicator in the
following positions:
1st position: 2
–
nd harmonic;
2nd position: crystal stat;
3rd position: cavity stat;
4th position: purifier current;
5th position: +27V;
6th position: +18V;
7th position: +12.6V;
8th position:
–
12.6V;
9th position: +5V;
10th position: 5
A;
11th position: 50
A;
12th position: 500
A.
Micro
–
ammeter
–
operating mode
indication of power supplies and
instrument separate units.
17
U
Button
–
ion pump voltage check.
Down
19
PUMP
Toggle switch
–
ion pump switch on.
20
AFC SUPPLY
Toggle switch
–
AFC unit switch on.
Down
21
HFO
Toggle switch
–
turns on HFO of
atomic hydrogen source.
Down
22
I
HFO
Button
–
I
HFO
check
23
5 MHz
Connectors
–
5 MHz signal outputs.
24
SYNCH. DET 1
Connector
–
crystal oscillator AFC
unit synchronous detector signal
output.
25
115/120/220/240V
50/400Hz
Plug
–
line frequency and voltage
selection.
Required line
frequency and
voltage
26
IF
Connector
–
19.6 MHz IF signal
output used only during adjustment,
instrument repair and Zeeman
frequency (f
z
) measurement.
Содержание CH1-76A
Страница 2: ...2 Fig 1 The CH1 76A Passive Hydrogen Maser ...
Страница 8: ...8 Fig 5 1 The Maser CHI 76A Simplified Block Diagram ...
Страница 9: ...9 Fig 5 2 The Maser CH1 76A Detail Block Diagram ...
Страница 12: ...12 Fig 6 1 The Maser CH1 76A Package Design ...
Страница 14: ...14 Fig 7 1 The Maser CH1 76A Front A and Rear B Panel ...
Страница 26: ...26 Fig 10 1 The Maser CH1 76A Accessory Kit ...