The  fastest  way  is  to  short  the  leads  to  auto-range  to  the  60
range,  then  press  the 

RANGE

  button  momentarily.  The  display

shows 

“Shrt”

. Keep shorting the leads for further 3 seconds until

the  display  shows  zero.  The  lead  resistance  is  then  temporarily
compensated.

W

The  compensation  value  stays  until  the  next  meter  power  reset,
and can be as high as 5

W

. If you need a compensation value that

is higher than that, Relative Zero mode is recommended.

Diode Test,      Capacitance Function

Press the 

SELECT

 button momentarily to toggle the functions. Last

selection will be saved as power up default for repeat measurement
convenience.

Test Voltage

SELECT

RANGE

INSULATION

TEST

REC

R

LOCK

REL

OFF

Hz

VFD

V

V

W

Hz

mV

mA

A

Hz

Hz

µ

A

INSULATION

50V...1000V

INSULATION

1000V CAT III

MAX

600V CAT IV

HBC FUSED

MAX 0.6A

HBC FUSED

MAX 10A

COM

A

JACK

JACK

 INS.

µA mA

 INS.

W

V

HOLD

20kHz ACV
Bandwidth

.

+

+

Test Voltage

SELECT

RANGE

INSULATION

TEST

REC

R

LOCK

REL

OFF

Hz

VFD

V

V

W

Hz

mV

mA

A

Hz

Hz

µ

A

INSULATION

50V...1000V

INSULATION

1000V CAT III

MAX

600V CAT IV

HBC FUSED

MAX 0.6A

HBC FUSED

MAX 10A

COM

A

JACK

JACK

 INS.

µA mA

 INS.

W

V

HOLD

20kHz ACV
Bandwidth

.

FORWARD BIAS

1

SELECT

+

Test Voltage

SELECT

RANGE

INSULATION

TEST

REC

R

LOCK

REL

OFF

Hz

VFD

V

V

W

Hz

mV

mA

A

Hz

Hz

µ

A

INSULATION

50V...1000V

INSULATION

1000V CAT III

MAX

600V CAT IV

HBC FUSED

MAX 0.6A

HBC FUSED

MAX 10A

COM

A

JACK

JACK

 INS.

µA mA

 INS.

W

V

HOLD

20kHz ACV
Bandwidth

.

REVERSE BIAS

+

Discharge capacitors before making any measurement. Large value
capacitors should be discharged through an appropriate resistance
load.

CAUTION :

®

18

Normal  forward  voltage  drop  (forward  biased)  for  a  good  silicon
diode  is  between  0.400V  to  0.900V.  A  reading  higher  than  that
indicates  a  leaky  diode  (defective).  A  zero  reading  indicates  a 
shorted diode (defective). An OL indicates an open diode (defective).
Reverse  the  test  leads  connections  (reverse  biased)  across  the 
diode. The digital display shows OL if the diode is good. Any other
readings indicate the diode is resistive or shorted (defective).

+Hz

DCmV, ACmV  functions

 :

Press  the 

SELECT

  button  momentarily  to  toggle  the  subject

functions.  Last  selection  will  be  saved  as  power  up default for
repeat measurement convenience.

mV

Hz

mV

Test Voltage

SELECT

RANGE

INSULATION

TEST

REC

R

LOCK

REL

OFF

Hz

VFD

V

V

W

Hz

mV

mA

A

Hz

Hz

µ

A

INSULATION

50V...1000V

INSULATION

1000V CAT III

MAX

600V CAT IV

HBC FUSED

MAX 0.6A

HBC FUSED

MAX 10A

COM

A

JACK

JACK

 INS.

µA mA

 INS.

W

V

HOLD

20kHz ACV
Bandwidth

0

6

.

1

SELECT

+

Test Voltage

SELECT

RANGE

INSULATION

TEST

REC

R

LOCK

REL

OFF

Hz

VFD

V

V

W

Hz

mV

mA

A

Hz

Hz

µ

A

INSULATION

50V...1000V

INSULATION

1000V CAT III

MAX

600V CAT IV

HBC FUSED

MAX 0.6A

HBC FUSED

MAX 10A

COM

A

JACK

JACK

 INS.

µA mA

 INS.

W

V

HOLD

20kHz ACV
Bandwidth

.

0

6

.

mV

Hz

+

Содержание KM 887

Страница 1: ...seconds on and 25 seconds off Dimension 208 L X 103 W X 64 5 H mm with holster Weight 635 gm with holster TRUE RMS DIGITAL INSULATION MULTIMETER MODEL KM 887 Preliminary Data D Chhaya My Documents Chhaya backup catlogs NEW CAT 2015 new arrival catalogs KM887 cdr An ISO 9001 2008 Company SPECIAL FEATURES All Specifications are subject to change without prior notice NEW Dual Display Hz readings Rela...

Страница 2: ... 1 20MW Add 2 operation temperature 35 C 600 0kW 100 W 6 000M 3 W 1 kW 1 rdg 3dgts 60 00M 4 W 10 kW 1 5 rdg 6dgts 5 6 600 0W 100 mW 0 2rdg 3dgts 0 3 rdg 2dgts AUDIBLE CONTINUITY TESTER Audible Threshold Between 20 and 350 W W Response Time 30ms approx Input Impedance 10MW 110pF nominal 1 Add 20d 80 of range 2 Unspecified 5 of range Range AC VOLTAGE Accuracy Resolution 50Hz 60Hz 0 7 rdg 4dgts 1 mV ...

Страница 3: ... 000M 30 00M 300 0M W W W 1000 V 3000MW 25 0GW 1mA 1MW 10 rdg 5dgts 1 5 rdg 5dgts 2 0 rdg 5dgts 550 0MW AC CURRENT Range Resolution Accuracy Burden Voltage 1 2 10A continuous upto ambient 35 C 15 mins on per 5 mins off 35 C 50 C 10A to 20A for 30 second on per 5 mins off 400mA continuous 400mA for 1 1 hours on per 20 minutes off 600 0 mA 60 00 mA 600 0 mA1 0 2mV mA 3mV mA 30mV A 6000 mA 6 000 A 10...

Страница 4: ...er Distance Meter Digital Lux Meter Network Cable Tester Power Factor Regulator Earth Resistance Tester Digital Panel Meters DC Power Supplies High Voltage Detector Calibrators Gas Analysers Frequency Counter Function Generator Phasing Sticks Battery Tester Waterproof Pen Testers Solar Power Meter EMF Detector Wood Paper Grain Moisture Meter Transistorised Electronic Analog Digital Insulation Resi...

Страница 5: ...R MODEL KM 887 TABLE OF CONTENTS TITLE PAGE NO I II Cenelec Directives Safety IV General Specifications III Special Features V Product Description Test Certificate IX Warranty VI Operation VII Maintenance VIII 01 03 04 05 14 15 25 27 28 ...

Страница 6: ...Part 2 Ed 2 0 and Part 10 Ed 2 0 These relevant parts of EN61557 are not covered by the ETL certification The accompanied test probe assembly meets UL IEC EN61010 031 Ed 1 1 to the same meter ratings or better IEC 61010 031 requires exposed conductive test probe tips to be 4mm for CAT III CAT IV ratings Refer to the category markings on your probe assemblies as well as on the add on accessories li...

Страница 7: ...ehold appliances portable tools similar equipment 04 IV GENERAL SPECIFICATIONS Display 3 5 6 digits 6 000 counts Polarity Automatic Update Rate 5 per second nominal 61 Segments Bar graph 40 per second max Operating Temperature 10 C to 50 C Relative Humidity Maximum relative humidity 90 for temperat ure up to 28 C decreasing linearly to 50 relative humidity at 50 C Pollution Degree 2 IP Rating IP40...

Страница 8: ...KB32 banana plug to type K socket plug adaptor BMH 01 magnetic hanger BMP 86x soft carrying pouch Special Features Record MAX MIN regular readings Relative Zero Display Hold LCD Backlight VFD V Hz readings Dual display Hz Readings High resolution 60 00mV 60 00 ranges Lock Test mode for Insulation resistance TM BeepJack audible visible input warning ELECTRICAL SPECIFICATIONS Accuracy is reading dig...

Страница 9: ...nge Resolution 1 Accuracy 10Hz 45Hz 45Hz 200Hz 0 1 0 1 600 0V 600 0V 4 0 rdg 5dgts 2 5 rdg 5dgts 08 200Hz 440Hz 0 1 600 0V 2 9 0 rdg 5dgts Input impedance 10MW 110pF nominal 1 Unspecified for fundamental frequency 440Hz 2 Accuracy linearly decreases from 2 5 5d 200Hz to 9 0 5d 440Hz Range Resolution Accuracy DC Voltage 0 01 60 00mV 0 2 rdg 3dgts Input impedance 10MW 110pF nominal 600 0mV 6 000V 60...

Страница 10: ...C 0 0 C to 50 0 C 50 0 C to 537 0 C 40 0 F to 32 0 F 32 0 F to 122 0 F 122 0 F to 999 0 F 2 2 C 1 2 C 1 3 6 F 4 F 1 3 6 F 1 Accuracies assume meter interior has the same temperature of the ambient isothermal stage for a correct junction voltage compensation Allow enough time to reach the isothermal stage for a significant change of ambient temperature It can take up to an hour for changes 5 C 2 Ty...

Страница 11: ...6 000A 10 00A 2 0 2mV µA 3mV mA 30mV A 0 1 1 0 01 0 1 0 001 0 01 Unspecified 12 1 400mA continuous 400mA for 1 1 hours on per 20 minutes off 2 10A continuous up to ambient 35 C 15 mins on per 5 mins off 35 C 50 C 10A to 20A for 30 seconds on per 5 mins off _ Hz Line Level Frequency Function Range Range Sensitivity Sine RMS 10Hz 100kHz 60mV 600mV 6V 60V 600V 1000V VFD 600V 600 A µ 6000 A µ 60mA 600...

Страница 12: ...your particular model for function availability 1 3 5 6 digits 6000 counts dual displays 2 Push buttons for special functions features 3 Selector to turn the Power On or Off and Select a function 4 5 C o m m o n G r o u n d reference Input Jack for all functions EXCEPT Insulation Resistance function 6 Input Jack for all functions EXCEPT A mA and µA functions 7 Input Jack for A function positive in...

Страница 13: ...e calibration on this range W 15 IV OPERATION CAUTION Before and after hazardous voltage measurements test the voltage function on a known source such as line voltage to determine proper meter functioning Hz Hz ACV VFD ACV Functions Press the SELECT button momentarily to toggle the subject functions Last selection will be saved as power up default for repeat Hz measurement convenience For ACV func...

Страница 14: ...0V 1000V INSULATION 1000V CAT III MAX 600V CAT IV HBC FUSED MAX 0 6A HBC FUSED MAX 10A COM A JACK JACK INS µA mA INS W V HOLD 20kHz ACV Bandwidth REVERSE BIAS Discharge capacitors before making any measurement Large value capacitors should be discharged through an appropriate resistance load CAUTION 18 Normal forward voltage drop forward biased for a good silicon diode is between 0 400V to 0 900V ...

Страница 15: ...ndwidth 0 6 A A Hz mA Hz µA Hz Test Voltage SELECT RANGE INSULATION TEST REC R LOCK REL OFF Hz VFD V V W Hz mV mA A Hz Hz µA INSULATION 50V 1000V INSULATION 1000V CAT III MAX 600V CAT IV HBC FUSED MAX 0 6A HBC FUSED MAX 10A COM A JACK JACK INS µA mA INS W V HOLD 20kHz ACV Bandwidth 0 6 Test Voltage SELECT RANGE INSULATION TEST REC R LOCK REL OFF Hz VFD V V W Hz mV mA A Hz Hz µA INSULATION 50V 1000...

Страница 16: ...Hz ACV Bandwidth 250V Test Voltage 1 Test Voltage SELECT RANGE INSULATION TEST REC R LOCK REL OFF Hz VFD V V W Hz mV mA A Hz Hz µA INSULATION 50V 1000V INSULATION 1000V CAT III MAX 600V CAT IV HBC FUSED MAX 0 6A HBC FUSED MAX 10A COM A JACK JACK INS µA mA INS W V HOLD 20kHz ACV Bandwidth 22 Default startup primary display reading is Allow enough time for a good measuring result After the ink is re...

Страница 17: ...ured reading that is beyond 300 counts in changes comparing to its former one Press the button momentarily to enable with LCD annuciator turned on Press momentarily again to disable 23 Test Voltage SELECT RANGE INSULATION TEST REC R LOCK REL OFF Hz VFD V V W Hz mV mA A Hz Hz µA INSULATION 50V 1000V INSULATION 1000V CAT III MAX 600V CAT IV HBC FUSED MAX 0 6A HBC FUSED MAX 10A COM A JACK JACK INS µA...

Страница 18: ...y switch OFF and then back on to resume MAINTENANCE WARNING To avoid electrical shock disconnect the meter from any circuit remove the test leads from the input jacks and turn OFF the meter before opening the case Do not operate with open case Install only the same type of fuse or equivalent Calibration Accuracy is specified for a period of one year after calibration Periodic calibration at interv...

Страница 19: ...CO authorized dealer This warranty does not apply for damaged Ic s fuses burnt PCB s disposable batteries carrying case test leads or to any product which in KUSAM MECO s opinion has been misused altered neglected contaminated or damaged by accident or abnormal conditions of operation or handling KUSAM MECO authorized dealer shall extend this warranty on new and unused products to end user custome...

Отзывы: