05
EMC :
Meets EN61326-(1997, 1998/A1), EN61000-4-2(1995), &
EN61000-4-3(1996).
In an RF field of 3V/m :
Capacitance function is not specified
AC 4.000V range : Total Accuracy = Specified Ac 700 dgts
AC 400.0
A range : Total Accuracy = Specified Ac 300 dgts
Other function ranges : Total Accuracy = Specified Ac 40 dgts
Performance above 3V/m is not specified.
Accessories :
Test lead (pair), Holster, Battery, User Manual & Carrying case.
DC VOLTAGE
400.0 mV
4.000 V
40.00 V
400.0 V
NMRR :
>50dB @ 50/60Hz
CMRR :
>120dB @ DC, 50/60Hz, Rs=1k
Input Impedance :
10M
, 30pF nominal
(1000M
for 400.0 mV range)
±(0.3%rdg + 4dgts)
±(0.5%rdg + 3dgts)
±(0.5%rdg + 3dgts)
±(0.5%rdg + 3dgts)
Accuracy
Resolution
Range
1000 V
±(1.0%rdg + 4dgts)
100
V
1 mV
10 mV
100 mV
1 V
06
ELECTRICAL SPECIFICATIONS :
Accuracy is ± (%readings number of digits) or otherwise
o
o
specified, at 23 C ± 5 C & less than 75% R.H.
Maximum Crest Factor <1.75:1 at full scale & <3.5:1 at half scale, &
with frequency components within the specified frequency
bandwidth for non-sinusoidal waveforms.
AC VOLTAGE
Accuracy
Resolution
Range
400.0 mV*
4.000 V
40.00 V
400.0 V
CMRR :
>60dB @ DC to 60Hz, Rs=1k
Input Impedance :
10M
, 30pF nominal
* Selection by RANGE Button manually, and is specified from
AC 10mV & up.
±(4.0%rdg + 5dgts)
±(1.5%rdg + 5dgts)
±(1.5%rdg + 5dgts)
±(1.5%rdg + 5dgts)
1000 V
±(4.0%rdg + 5dgts)
50Hz ~ 500Hz
1 V
100
V
1 mV
10 mV
100 mV