II) ELECTRICAL SIGNS
Extremely dangerous! Operators shall strictly ad here
to safety rules; otherwise there would be dangers
of electric shock to cause personal injuries or casualties
Dangerous! Operators shall strictly adhere safety rules;
otherwise there would be dangers of electric shock to
cause personal injuries or casualties.
!
Warning! Operators shall strictly keep
safety rules;
otherwise personal injuries or equipment damages
might be caused.
ad here the
!
Double insulation
AC
DC
III) Range and Accuracy
Function
Range
Resolution
Accuracy
0.1A
DC current 0.0A~600A DC ±2%rdg±3dgts
AC current
0.1A
0.0A~600A AC ±2%rdg±3dgts
IV) ELECTRICAL SPECIFICATIONS
Function
Power Supply
Test Mode
Display Mode
Dimension
LCD Dimension
Sampling Rate
Frequency
Polarity
Indication
Measurement of AC/DC current,
Peak value hold, on-line monitoring
Standard 9V battery
Clip-on CT, integral mode
Four digits LCD display
LWH : 175mm x 70mm x 38mm
35mm x 21.5mm
display domain : 32mm x 15mm
2 times/s
50/60Hz Automatically
DC current auto indentified and
display " "
Clamp Size
Test Position
Range Shift
Line Voltage
RS232 Interface
Com-Configure
Data Memory
30mm x 35mm
Tested wire in the jaw center; Error
may increase by 1.5%rdg max if
deviate the center
Automatically
AC600V
Data Stored in the memory of the
meter via RS232 uploaded to PC,
or on-line monitoring
Baud rate : 9600, data bit :8, stop bit :1
99units, "
FULL"
blinks when the
memory is full
Peak Hold
Reading Hold
Out of Range
Auto Power-off
Battery Voltage
Push "
HOLD"
without release, the
meter will show the peak value
"DH"
indicating the reading is hold
"OL"
indicating the current is out of range
5 Minutes after power on, it will power
off automatically to lower the power
consumption
Indicating the battery voltage is
lower then 7.2V. Then the battery
has to be changed
Weight
Consumption
Temperature
and Humidity
Max error refer
to environment
Insulating
Strength
180g (including the battery)
10mW
Working : -10°C~50°C, below 80% RH
Storage : -10°C~60°C, below 70% RH
10°C~0°C, 40°C~50°C,
Error will increase max by 1% rdg
AC2kV/rms (between the alloy of
the clamp and the housing)