5.4 Measurement Spotlight LED
The measurement Spotlight LED helps to
quickly find mA signal wires. Press the
Spotlight button to turn the Spotlight on or off.
5.5 Display HOLD
To avoid electric shock, when Display HOLD is
activated, the display will not change when a
different current is applied.
Press the HOLD button to activate Display
Hold mode.
The symbol “DH” are displayed and the Meter
freezes the display. To exit & return to normal
operation, press the HOLD button a second
time.
Warning
Warning
The Clampmeter is not for use on non-insulated
conductors.
Measurements can be taken with the clamp
approach into any position.
For accurate measurements :
?
Always zero the Meter prior to taking
measurements.
?
Zero the Meter as close to the measurement
source position as possible.
?
Make sure the clamp is free of contamination.
6
KUSAM-MECO July 2011
®
6. MEASUREMENT
?
With the clamp disconnected from any
conductor, press the power switch to turn the
meter ON. Then, you might wait for about 5
seconds.
?
Clamp the jaw around the conductor under
test. The meter displays the measured
conductor current.
?
A positive reading indicates current flowing in
the direction of the arrow on the clamp.
?
A negative reading indicates current flowing in
the opposite direction of the arrow.
?
Do not clamp more than one wire. Currents
cancel and no results are returned.
The small secondary display shows the reading
in terms of percentage span.
Note :
To reduce magnetic influences, zero the Meter in
the same position or jaw direction that is used for
the measurement.