Power ON / OFF For Battery-Operated Functions

The  battery-operated  portion  of  the  tester  turns  on
automatically when the two test probes are shorted
together,  or  significant  operating  voltage  level  is
detected. See specifications for power-on threshold
details.  Where  available,  momentary  press  the
Resistance,  Capacitance  or  Backlight  button  can
also manually turn on the tester. The tester turns off
automatically  after  no  further  significant  testing  or
push-button  activities  for  approximately  8  seconds
for  Resistance  &  Capacitance  functions  or  16
seconds for others.

To avoid electrical shock, disconnect the tester from 
any circuit and remove the test probes from the input 
jacks  before  opening  the  case  and  /  or  the  battery 
access  door.  Do  not  use  the  tester  with  open  case 
and  /  or  the  battery  access  door.  Do  not  attempt  to 
repair this unit.  It contains no user-serviceable parts. 
Unauthorized  persons  shall  not  disassemble  the 
tester.

WARNING

 MAINTENANCE

Periodically wipe the case with a damp cloth and mild
detergent;  do  not  use  abrasives  or  solvents.  The 
tester should be kept dry and clean. If the tester is not 
to be used for periods of longer than 60 days, remove 
the batteries and store them separately.

Cleaning and Storage

Trouble Shooting

22

KUSAM-MECO                                                                              Feb  2018

®

If the tester fails to operate, check batteries, probes
etc.,  and  replace  as  necessary.    Double  check
operating  procedure  as  described  in  this  user’s
manual.

If the tester voltage-continuity input has subjected to
high voltage transient (mostly caused by lighting or
switching  surge  to  your  system)  by  accident  or
abnormal conditions of operation, the series fusible
resistors will be blown off (become high impedance) 
like  fuses  to protect  the user and the tester. Most
measuring  functions  through  this  input  will  then  be
open  circuit.  The  series  fusible  resistors  and  the
spark  gaps  should  then  be  replaced  by  qualified
technician.

Low-Battery Indication

WARNING

 

 Always power-on the tester and check 

for low-battery status before using battery-operated 
functions. Replace low-batteries ASAP to maintain 
tester accuracy and functionality.

EF and Continuity LEDs flashing simultaneously to 
Indicate Low-Battery. 

(KM 67)

The LCD annunciator      turns on to Indicate Low-
Battery. 

(KM 69)

Loosen the 2 screws from the battery access door 
of the case bottom. Lift the battery access door up.  
Replace the batteries. Re-fasten the screws.

Battery Replacement

The tester uses 1.5V AAA Size (NEDA 24A OR IEC 
LR03) Alkaline Battery x 2

Содержание 181041412

Страница 1: ...C Clamp Adaptor AC DC Current Adaptor Transistorised Electronic Analog Digital Insulation Resistance Testers upto 10 KV Digital Sound Level Meter Sound Level Calibrator Digital contact Non contact Type Tachometer Digital Non contact infrared Thermometer Thermo Hygrometer Thermo Anemometer Wood Paper Moisture Meter Distance Meter Digital Hand Held Temperature Indicators Digital Lux Meter Network Ca...

Страница 2: ...ical Specification 08 Product Description 11 Operation 12 Maintenance 21 Test Certificate 23 Warranty 24 Electrical Symbols 03 KUSAM MECO Feb 2018 VOLTAGE TESTER MODEL KM 66 KM 69 EF EF EF ELV 690 1000 400 230 120 50 24 12 V V V RCD 30mA 230V EF EF EF ELV 690 1000 400 230 120 50 24 12 V V V RCD 1 8 8 8 H AUTO V L R ...

Страница 3: ...43 3 2014 It is however not intended to 02 KUSAM MECO Feb 2018 be used in wet locations CAUTION Disconnect the test leads from the test points before changing functions The voltages marked on the tester are nominal voltages or nominal voltage ranges The tester shall only be used on installations with the specified nominal voltages or nominal voltage ranges The different indicating signals of the v...

Страница 4: ...such as stationary motors with permanent connection to the fixed installation 3 Measurement Category II is applicable to test and measuring circuits connected directly to utilization points socket outlets and similar points of the low voltage MAINS installation Examples are measurements on MAINS CIRCUITS of household appliances portable tools and similar equipment 04 KUSAM MECO Feb 2018 SAFETY ADV...

Страница 5: ...ating voltage on the part to be tested and to conclude that the voltage indicated by the voltage detector is an interference voltage 2 CENELEC DIRECTIVES LCD indication KM 69 AC Detection Shaker KM 69 RCD Leakage Path KM 69 RST Phase Rotation Detection KM 69 EF Detection SPECIAL FEATURES 06 KUSAM MECO Feb 2018 SPECIFICATION GENERAL SPECIFICATION LCD Display KM 69 Display Update Rate KM 69 5 per se...

Страница 6: ...nary red probe and detachable leaded black probe 08 KUSAM MECO Feb 2018 ELECTRICAL SPECIFICATIONS DC AC Voltage LED Accuracy is given as of reading digits number of digits or otherwise specified 23 C 5 C Maximum Crest Factor 2 1 at full scale 4 1 at half scale and with frequency components fall within the tester specified frequency bandwidth for non sinusoidal waveforms LED Threshold ON 85 of Volt...

Страница 7: ...ection 10 KUSAM MECO Feb 2018 flashing rate of Continuity LED accompanied with beep sound KM 69 also display successive bar graph segments proportional to the field strength on LCD Typical Non Contact EF Detection values are shown in the following table LCD Bar Graph Indication KM 69 15V to 55V 30V to 95V 55V to 170V Above 120V KM 66 KM 69 80V to 250V 150V to 450V 300V to 700V Above 500V Typical N...

Страница 8: ...shortly before and after a test If indication of one or more steps fails or if no functioning is indicated the tester shall no longer be used Before using the Audible Continuity EF Detection features at locations with a high backgroud noise level it shall be determined whether the audible signal is perceptible The audible indication is for information only do not rely on it especially in high back...

Страница 9: ... from the system under test without the need of internal battery supply battery less 14 KUSAM MECO Feb 2018 WARNING When activated the tester actually connects to a low impedance load of nominal 6 1kW to draw a leakage current of 30mA typical at 230VAC to trip RCD circuit breakers In the event of a full continuous 5 second activation interval the protection PTC will heat up and affect the load cur...

Страница 10: ...tor R turns on it indicates that a Right phase rotation is being detected The Red nd probe L2 connection is then the 2 Phase S The rd remaining unconnected connection is the 3 Phase T Swap the two probe connections to get reversed annunciator L as for proper functioning verification Note This function works only on Y connection 3P4W systems It does not work on Δ connection 3P3W systems that do not...

Страница 11: ...distinguish between live or earth connections As illustrated press and hold the EF button to enter and remain at EF Detection feature The EF LED turns on Signal strength is indicated by the flashing rate of Continuity LED accompanied with beep sound KM 69 further display EF when they are ready and display signal strength as successive bar graph segments on the LCD S T R L1 L2 EF RCD 30m A 230 V EF...

Страница 12: ...n 20 KUSAM MECO Feb 2018 During power on KM 69 LCD further display _I _ for open circuits and _ _ _ for complete circuits Audible Continuity is convenient for checking wiring connections and operation of switches Press button momentarily to toggle on off the backlight flashlight Backlight Flashlight KM 69 EF RCD 30mA 230V EF EF ELV 690 1000 400 230 120 50 24 12 RCD V V V Live Ground ON flash H EF ...

Страница 13: ... for periods of longer than 60 days remove the batteries and store them separately Cleaning and Storage Trouble Shooting 22 KUSAM MECO Feb 2018 If the tester fails to operate check batteries probes etc and replace as necessary Double check operating procedure as described in this user s manual If the tester voltage continuity input has subjected to high voltage transient mostly caused by lighting ...

Страница 14: ...KUSAM MECO authorized dealer This warranty does not apply for damaged Ic s fuses burnt PCB s disposable batteries carrying case test leads or to any product which in KUSAM MECO s opinion has been misused altered neglected contaminated or damaged by accident or abnormal conditions of operation or handling KUSAM MECO authorized dealer shall extend this warranty on new and unused products to end user...

Отзывы: