![Kurt J. Lesker KJLC XCG Скачать руководство пользователя страница 22](http://html1.mh-extra.com/html/kurt-j-lesker/kjlc-xcg/kjlc-xcg_operating-manual_2011293022.webp)
22
tkna80e1 (2018-08)
Caution
Dirt sensitive area
Touching the product or parts thereof with bare
hands increases the desorption rate.
Always wear clean, lint-free gloves and use clean
tools when working in this area.
Vent the vacuum system.
Put the gauge out of operation.
Disconnect the sensor cable.
Remove the gauge from the vacuum system and install the
protective lid.
6
Maintenance, Repair
The product requires no maintenance.
Gauge failures due to contamination are not covered by
the warranty.
We recommend checking the zero at regular intervals.
Kurt J. Lesker Company assumes no liability and the warranty
becomes null and void if any repair work is carried out by the
end-user or third parties.