background image

Washing lymphocytes/ RS-110L rotor

Turn the “HLA-SERO selector switch” to the 

HLA side.

Washing erythrocytes/ Seromatic rotor

Turn the “HLA-SERO selector switch” to the 

SERO side.

    Usage

Separating and washing lymphocytes

Separating and purifying of cell culture

Washing blood corpuscles (coombs test), blood test etc.

Rotor for washing lymphocytes

Rotor

name

Push

button

Speed

rpm

RCF

xg

Time

(changeable range)

Usage

RS-110L

1

4,700

2,000

180 sec

(30~300 sec)

Separating 

lymphocytes and cell 

culture

2

3,300

1,000

3 sec

Removing platelet 

(thrombin preparation)

3

3,300

1,000

60 sec

(30~120 sec) Washing lymphocytes

Rotor for washing erythrocytes

Rotor

name

Push

button

Speed

rpm

RCF

xg

Time

(changeable range)

Usage

Seromatic

1

2,200

500

60 sec

(30~120 sec)

Blood test, observing 

cohesive action of 

blood corpuscles

2

3,100

1,000

15 sec

(5~25 sec)

Cross-match test, 

antiglobulin test

3

3,100

1,000

60 sec

(30~120 sec)

Washing blood 

corpuscles, extraction 

of lymph and blood 

plasma from 

dissociate solution

*1

*3

*2

“THROMBIN” knob

“HLA-SERO” selector switch

“SELECTOR” switch

Rotor is optional.

Rotor is optional.

KA-2200

*1  12x0.25-1mL tubes can be used.

*2  Time is changeable for 2-6 sec by time setting knob on front panel.

*3  12x glass tubes (length 65-80mm, diam. 10-13mm) can be used.

Rotor has prescribed years of operation, after that the rotor must be replaced.

For more details, please contact us.

Отзывы: