Washing lymphocytes/ RS-110L rotor
Turn the “HLA-SERO selector switch” to the
HLA side.
Washing erythrocytes/ Seromatic rotor
Turn the “HLA-SERO selector switch” to the
SERO side.
Usage
●
Separating and washing lymphocytes
●
Separating and purifying of cell culture
●
Washing blood corpuscles (coombs test), blood test etc.
Rotor for washing lymphocytes
Rotor
name
Push
button
Speed
rpm
RCF
xg
Time
(changeable range)
Usage
RS-110L
1
4,700
2,000
180 sec
(30~300 sec)
Separating
lymphocytes and cell
culture
2
3,300
1,000
3 sec
Removing platelet
(thrombin preparation)
3
3,300
1,000
60 sec
(30~120 sec) Washing lymphocytes
Rotor for washing erythrocytes
Rotor
name
Push
button
Speed
rpm
RCF
xg
Time
(changeable range)
Usage
Seromatic
1
2,200
500
60 sec
(30~120 sec)
Blood test, observing
cohesive action of
blood corpuscles
2
3,100
1,000
15 sec
(5~25 sec)
Cross-match test,
antiglobulin test
3
3,100
1,000
60 sec
(30~120 sec)
Washing blood
corpuscles, extraction
of lymph and blood
plasma from
dissociate solution
*1
*3
*2
“THROMBIN” knob
“HLA-SERO” selector switch
“SELECTOR” switch
Rotor is optional.
Rotor is optional.
KA-2200
*1 12x0.25-1mL tubes can be used.
*2 Time is changeable for 2-6 sec by time setting knob on front panel.
*3 12x glass tubes (length 65-80mm, diam. 10-13mm) can be used.
Rotor has prescribed years of operation, after that the rotor must be replaced.
For more details, please contact us.