OPERATION
5
23
SMARTPAT PH 8530
www.krohne.com
03/2016 - 4002544803 - MA SMARTPAT PH 8530 R03 en
5.1 Calibration
5.1.1 Calibration with PACTware
TM
Manual hold
•
Start the function calibration
calibration
calibration
calibration in menu mode Quick Setup
Quick Setup
Quick Setup
Quick Setup or Setup
Setup
Setup
Setup.
•
Activate the function manual hold
manual hold
manual hold
manual hold to avoid an alarm.
•
Select yes
yes
yes
yes.
•
Press Next
Next
Next
Next to proceed.
Temperature compensation
•
Set the temperature compensation
temperature compensation
temperature compensation
temperature compensation for calibration.
•
Select between automatic
automatic
automatic
automatic or manual
manual
manual
manual.
i
If you have selected manual temperature compensation
manual temperature compensation
manual temperature compensation
manual temperature compensation, please enter the temperature of
the pH buffer solution. Default setting: 25
°
C / 77
°
F.
•
Press Next
Next
Next
Next to proceed.
Start calibration procedure
•
Dip the sensor into the first pH buffer solution
1
and wait till the value is stable.
•
Press Next
Next
Next
Next to proceed.
i
The measurement of pH buffer solution 1
pH buffer solution 1
pH buffer solution 1
pH buffer solution 1 starts and the values pH and temperature appear.
After approx. 10 seconds the following values appear:
pH value
temperature value
identified buffer value
•
Press Next
Next
Next
Next to confirm the identified pH buffer solution. If necessary change the value of the pH
buffer solution manually. If the sensor does not recognise the pH buffer solution the message
buffer not recognised
buffer not recognised
buffer not recognised
buffer not recognised appears.
•
Press Next
Next
Next
Next to proceed.
i
The value 0 for the pH buffer solution appears. Enter the value of the pH buffer solution
manually.
•
Press Next
Next
Next
Next to proceed with the calibration procedure.
•
Clean the sensor with demineralised water
2
.
Figure 5-1: Calibration procedure