3 Calibration 

Calibration with PACTware

TM

 (download free of charge on KROHNE.com) e.g. with OPTIBRIDGE

 

Manual hold 

select Menu mode 

Quick Setup

 or 

Setup

 / start 

calibration

 / activate 

manual hold

 / 

select 

yes

 / press 

Enter

 

Start calibration 
procedure 

Set value of ORP solution; Default setting: 468 mV

 / press 

Next / 

Dip the sensor into 

the ORP solution and wait till the value is stable 

. / press 

Next

 

 

 

 

 

 

 

 

 

 

After successful 
calibration  

the following values appear: 

 

Old ORP value [mV], ORP solution [mV], Electrode voltage [mV], New offset 
[mV], New ORP value [mV]

 

press 

Next

 / set calibration date DD-MM-YYYY / press 

Next

 / 

Save values?

 / 

select 

yes

 / press 

Next 

/ select 

no

 to deactivate 

manual hold

 / press 

Next

 to leave 

calibration menu 

After unsuccessful 
calibration 

the message 

Warning! Offset out of range!

 appears behind the value 

New offset

 

 

Calibration with HART Handheld 475 FIELD COMMUNICATOR

 

Manual hold 

select Menu mode 

Quick Setup

 or 

Setup

 / start 

calibration

 / activate 

manual 

hold

 / select 

yes

 / press 

Enter

 

Start calibration 
procedure 

Set value of ORP solution; Default setting: 468 mV

. Press key 

>

 to enter submenu 

to 

change default setting.

 / press 

Enter

 and 

Ok / 

Dip the sensor into the ORP 

solution and wait till the value is stable 

. / press 

Ok

 

After successful 
calibration  

the following values appear: 

 

Old ORP value [mV], ORP solution [mV], Electrode voltage [mV], New offset 
[mV], New ORP value [mV]

 

Press 

Ok

 / set calibration date 

DD-MM-YYYY

 with key 

>

 / press 

Enter

 / press 

Ok

 / 

Save values?

 / select 

yes

 / press 

Enter

 / select 

no

 to deactivate 

manual hold

 / 

press 

Enter

 to leave calibration menu 

After unsuccessful 
calibration 

message 

Warning!!!

 

Offset out of range!

 appears behind the value 

New offset

 

 

 

 

 

 

 

Calibration with SMARTMAC 200 W

 

Manual hold 

select Menu mode 

Quick Setup

 or 

Setup

 / start ORP 

calibration

 / press > / 

Manual hold?

 / select 

yes

 / press 

Enter 

twice 

Start calibration 
procedure 

Set ORP solution

 e.g. 

220 mV 

/

 press >

 /

 

enter temperature for calibration

 

solution / wait for stable value

 and press 

Enter 

Please wait

 /  

After successful 
calibration  

the following 

Calibr.values

 appear:  

Offset [mV], ORP [mV]

 

Press 

Enter

 / 

Store cal. values?

 / select 

yes

 / press 

Enter

 twice / select 

Set clock

 

with 

>

 / set calibration date 

YYYY-MM-DD

 and time 

HH:MM / 

press 

enter / 

select 

Manual hold?

 With 

> / 

select

 no / 

press

 Enter

 twice / 

ORP calibration sucessful

 

/ press 

Enter

 several times to leave the calibration 