 DEVICE DESCRIPTION 

2

7

SMARTPAT COND 5200

www.krohne.com

03/2022 - 4004007703 - MA SMARTPAT COND 5200 R03 en

2.3  Flammable products

Atmospheric conditions:

The standard atmospheric conditions according to IEC 60079-0 under which it may be assumed 

that Ex equipment can be operated are:

Temperature: -20...+60°C / -4...+140°F

Pressure: 80...110 kPa (0.8...1.1 bar) / 11.6...15.9 psi

Air with normal oxygen content, typically 21%v/v

Ex equipment must be tested and certified for the intended operating temperature range (e.g. for 

ambient temperature range -10...+80°C / 14...+176°F).

Ex equipment operating outside the specified above standard atmospheric conditions is 

permitted only via risk assessment of the user and additional measures, as necessary.

Operating conditions:

The sensor part of SMARTPAT COND 5200 may operate outside the standard atmospheric 

pressure range, which means that explosion protection may not be applicable for the sensor part 

(piping).

2.4  Device category

The sensor is designed in category II 1 G for use in zone 0.

For more information refer to chapter "Flammable products".

DANGER!

Operation with flammable products is only permitted as long as no explosive fuel/air mixture 

builds up inside of the piping at the same time the atmospheric conditions are exceeded.

The operator is responsible to ensure that the sensor is operated safely in terms of temperature 

and pressure, as well as the medium used. In case of operation with flammable media the 

sensor must be included in the periodic pressure tests of the plant!

The sensor housing is made from non-conductive PVDF. In order to avoid danger from 

electrostatic charge, the minimum conductivity of the media must be 10 nS/cm.

INFORMATION!

Definition of zone 0 according to EN 1127-1:

An area in which an explosive atmosphere, as a result of the mixture of flammable substances in 

the form of gas, vapours or mist with air, is continuously or for long periods or frequently 

present.

Содержание SMARTPAT COND 5200

Страница 1: ...Digital conductivity sensor Category II 1 G EPL Ga SMARTPAT COND 5200 Supplementary Instructions KROHNE 03 2022 4004007703 MA SMARTPAT COND 5200 R03 en AD ...

Страница 2: ... 6 2 3 Flammable products 7 2 4 Device category 7 2 5 Protection type for gas hazardous areas 8 2 6 Temperature classes 8 2 7 Electrical data 9 3 Installation 10 3 1 Installation 10 3 2 Specific conditions of use 10 4 Electrical connections 11 4 1 General notes 11 4 2 Power supply 11 4 3 Inputs outputs 11 4 4 Grounding and equipotential bonding 12 5 Operation 13 5 1 Starting up for the first time ...

Страница 3: ...eclaration of conformity on his own responsibility conformity with the protection goals of directive 2014 34 EU for use in hazardous areas with gas Conformity with harmonised standards was checked by a notified body in accordance with EN IEC 60079 0 2018 and EN 60079 11 2012 The EC declaration of conformity is based on the EC type examination certificate You can download the EC type examination ce...

Страница 4: ...h EN IEC 60079 0 2018 and EN 60079 11 2012 by an UK Approved Body The UK declaration of conformity is based on the UK type examination certificate issued by CSA Group If necessary the UK type examination certificate can be downloaded from the manufacturer s website 1 5 Safety instructions Assembly installation start up and maintenance may only be performed by personnel trained in explosion protect...

Страница 5: ...e of the sensor is the measurement of conductivity in conductive liquids It has a connection for a 4 20 mA current output with HART communication The connection is designed as a plug connector and at the same time supplies the sensor with power 1 VP 2 0 connector 2 PEEK housing 3 G3 4 A thread male or 3 4 14 NPT male 4 Electrodes graphite PES ...

Страница 6: ...urer 2 Device name 3 Cell constant 4 TAG number 5 Order code 6 Serial number D14xxxxxxx Year of manufacture 2014 7 Ex data acc to KIWA 14ATEX0002 X or IECEx KIWA 14 0004X or CSAE 21UKEX2617X 8 Manufacturing date Ingress protection 9 Observe the installation and operating instructions SMARTPAT COND 5200 KIWA 14ATEX0002 X Ex ia IIC T6 T4 Ga SN D14xxxxxxx 0102 VGS C 1234567 II1G TAG ABCD1234 c 1 ER 1...

Страница 7: ...losion protection may not be applicable for the sensor part piping 2 4 Device category The sensor is designed in category II 1 G for use in zone 0 For more information refer to chapter Flammable products DANGER Operation with flammable products is only permitted as long as no explosive fuel air mixture builds up inside of the piping at the same time the atmospheric conditions are exceeded The oper...

Страница 8: ... heat radiation e g direct sunlight adjoining hot parts Insulation does not obstruct the ventilation of the sensor Components of the Ex marking and their definition II Group II explosion protection 1 Equipment category 1 G Gas explosion protection Ex ia Intrinsically safe equipment protection level ia IIC Gas groups IIA IIB IIC approved T6 T4 Temperature class range T6 T4 Ga EPL Ga approved suitab...

Страница 9: ...mA Pi 1 0 W Li 10 µH Ci 5 nF CAUTION Capacity and inductance of the connecting cable have to be considered INFORMATION For information about rated data and the HART interface consult the standard documentation The internal intrinsically safe electrode circuit is limited to the following maximum values Uo 7 8 V Io 15 5 mA Po 30 5 mW The connections for the intrinsically safe 4 20 mA sensor circuit ...

Страница 10: ...may only be cleaned with a wet cloth DANGER The manufacturer is not liable for any damage or injuries resulting from improper use or use other than the intended purpose This applies in particular to hazards due to insufficient corrosion resistance and suitability of the materials in contact with product DANGER Installation and setup must be carried out according to the applicable installation stan...

Страница 11: ... explosive area or carefully insulate them from each other and from ground test voltage 500Veff 4 2 Power supply The sensor does not require a separate power supply The required supply for the electronics is provided via the 4 20 mA current output 4 3 Inputs outputs The connector pin assignment of the sensor is described in the standard documentation The signal circuits of the sensor may only be c...

Страница 12: ...IEC 60079 14 When using junction box type SJB 200 W Ex the shield of the sensor cable must be connected to the shield connector provided This will ensure sufficient electrostatic grounding and at the same time prevent unacceptably high equalizing currents INFORMATION The connections for the intrinsically safe 4 20 mA sensor circuit are galvanically isolated from both the electrode circuit and the ...

Страница 13: ...able them during operation Include the sensor in the periodic pressure tests of the plant when operating with flammable products 5 3 Electrostatic charge In order to avoid ignition hazards due to electrostatic charge the sensors may not be used in areas where processes that generate strong charges mechanical friction and cutting processes spraying of electrons e g in the vicinity of electrostatic ...

Страница 14: ...smantling the electrical connecting lines of the device ensure that all the cables leading to the sensor are de energized themselves and to the reference potential of the explosive area All the relevant process pipes or containers must be depressurized before dismounting the sensor If the process pipes or the container contain products that are harmful to the environment or to health implement the...

Страница 15: ...NOTES 7 15 SMARTPAT COND 5200 www krohne com 03 2022 4004007703 MA SMARTPAT COND 5200 R03 en ...

Страница 16: ...solutions Engineering commissioning calibration maintenance and training services Head Office KROHNE Messtechnik GmbH Ludwig Krohne Str 5 47058 Duisburg Germany Tel 49 203 301 0 Fax 49 203 301 10389 info krohne de KROHNE 03 2022 4004007703 MA SMARTPAT COND 5200 R03 en AD Subject to change without notice The current list of all KROHNE contacts and addresses can be found at www krohne com ...

Отзывы: