OPERATION
6
49
OPTIWAVE 1010
www.krohne.com
10/2016 - 4003537403 - MA OPTIWAVE 1010 R03 en
6.7 Device settings
6.7.1 Password protection for device settings
To change the device settings in the DTM, it is necessary to enter the correct password and log in
as a "Supervisor". If you do not log in, you can only read the device settings.
Procedure: How to log in as a supervisor
•
(a) Double click on the "OPTIWAVE 1010" element in the project structure (Project pane) or (b)
Right click on "OPTIWAVE 1010" element in the project structure (Project pane) and select
"Parameter" in the drop-down list box.
•
Click on "Login/Logout".
•
Click on the Select action
Select action
Select action
Select action menu and set it to "Login as Supervisor".
•
Enter the password (the default password is 123412
123412
123412
123412).
•
Push the "Execute" button.
•
The device settings are unlocked.
i
The device settings are unlocked. You can change the device settings. End of the procedure.
Figure 6-11: Login/Logout menu
Procedure: How to change the supervisor password
Figure 6-12: Procedure: How to change the supervisor password