 DEVICE DESCRIPTION 

2

7

OPTISWIRL 5080

www.krohne.com

07/2020 - 4008054001 - AD OPTISW5080 Exd Gb-Db en R01

2.3  Safety specifications

2.4  Flammable products

Atmospheric conditions:

Atmospheric conditions:

Atmospheric conditions:

Atmospheric conditions:

The standard atmospheric conditions under which it may be assumed that Ex equipment can be 
operated are:

Temperature: -20...+60

°

C / -4...+140

°

F

Pressure: 80...110 kPa (0.8...1.1 bar) / 11.6...15.9 psi

Air with normal oxygen content, typically 21%v/v

Ex equipment operating outside the standard temperature range must be tested and certified 
(e.g. for ambient temperature range -40...+65

°

C / -40...+149

°

F).

Ex equipment operating outside the standard atmospheric pressure range and standard oxygen 
content is not permitted.

Operating conditions:

Operating conditions:

Operating conditions:

Operating conditions:

Vortex flowmeters operate outside the standard atmospheric pressure range, which means that 
explosion protection, regardless of the zone assignment, is fundamentally not applicable for the 
measuring unit (piping).

Type of protection

Hazardous area classification

Gas environment

Gas environment

Gas environment

Gas environment

Equipment protection by intrinsic safety

Equipment protection by intrinsic safety

Equipment protection by intrinsic safety

Equipment protection by intrinsic safety

(remote flow sensor only):

(remote flow sensor only):

(remote flow sensor only):

(remote flow sensor only):

ATEX category remote flow sensor: II 1 G

Equipment protection level (EPL) remote flow sensor: Ga

Ex marking remote flow sensor: Ex ia IIC T4 Ga

Equipment protection by flameproof enclosure:

Equipment protection by flameproof enclosure:

Equipment protection by flameproof enclosure:

Equipment protection by flameproof enclosure:

ATEX category compact version and remote signal converter: II 2/1 (1) G

Equipment protection level (EPL) compact version and remote signal 

converter: Gb

Ex marking compact version: Ex db [ia Ga] ia IIC T4 Gb 

Ex marking remote signal converter: Ex db [ia Ga] IIC T4 Gb

Zone 0 (remote flow sensor only)

Zone 1 or zone 2

Gas group IIA, IIB or IIC

Temperature class range T1-T4

Dust enviroment

Dust enviroment

Dust enviroment

Dust enviroment

Dust ignition protection by enclosure:

Dust ignition protection by enclosure:

Dust ignition protection by enclosure:

Dust ignition protection by enclosure:

ATEX category: II 2 D

Equipment protection level (EPL): Db

Ex marking remote flow sensor: Ex tb IIIC T103

°

C Db

Ex marking compact version and remote signal converter:

Ex tb IIIC T85

°

C Db

Zone 21 or zone 22

Dust group IIIA, IIIB or IIIC

Maximum surface temperature 103

°

C / 

217

°

F or 85

°

C / 185

°

F at process and 

ambient temperature 80

°

C / 176

°

F

Table 2-1: Safety specifications

CAUTION!

Operation with flammable products is only permitted as long as no explosive fuel/air mixture 
builds up inside of the piping at the same time the atmospheric conditions are exceeded.

The operator is responsible to ensure that the flowmeter is operated safely in terms of the 
temperature and pressure of the products used. In case of operation with flammable products 
the measuring units must be included in the periodic pressure tests of the piping.

Содержание OPTISWIRL 5080

Страница 1: ...2 D Equipment protection level Equipment protection level Equipment protection level Equipment protection level Ga Gb and Db OPTISWIRL 5080 OPTISWIRL 5080 OPTISWIRL 5080 OPTISWIRL 5080 Supplementary...

Страница 2: ...evice description 5 2 2 Marking 5 2 3 Safety specifications 7 2 4 Flammable products 7 2 5 Ambient temperature temperature classes 8 2 6 Electrical data 10 3 Installation 11 3 1 Mounting 11 3 2 Specia...

Страница 3: ...tion contains only the data relevant to explosion protection The technical details given in the manual for the non explosion protected versions remain unchanged unless they will be excluded or replace...

Страница 4: ...vice which have been listed in these instructions If needed the IEC certificate can be downloaded from the manufacturer s website 1 4 Safety instructions If these instructions are not followed there i...

Страница 5: ...2 1 Example of a nameplate for the compact version 1 Model number and serial number 2 Marking according to the notified body 3 Note for disposal 4 Manufacturer logo 5 IP rating for enclosure 6 Manufa...

Страница 6: ...intrinsic safety Do not open when energized Failure to follow these instructions can result in death or serious injury 47058 DUISBURG GERMANY ST S N ORIGIN SUPPLY MWP 100 F 38 C MAX AMB TEMP 176 F 80...

Страница 7: ...on level EPL remote flow sensor Ga Ex marking remote flow sensor Ex ia IIC T4 Ga Equipment protection by flameproof enclosure Equipment protection by flameproof enclosure Equipment protection by flame...

Страница 8: ...mbient temperature range Insulation must be limited to the piping Unobstructed ventilation of the signal converter and the remote flow sensor connection box must be ensured Temperature class Permissib...

Страница 9: ...e operator has to ensure the temperature limit by temperature measurements at the housing reference point For high temperature applications positioning the electronics housing to the side or bottom an...

Страница 10: ...must be disconnected to shorten the cable run the cable through conduit or for some other reason you must disconnect the cable at the sensor junction box end You cannot disconnect the cable at the si...

Страница 11: ...is clearly visible it can be operated from a location with secure footing Important notes for mounting Important notes for mounting Important notes for mounting Important notes for mounting Ensure ar...

Страница 12: ...s with a tensile strength of at least 240 N mm Devices marked as category 1 G or EPL Ga Devices marked as category 1 G or EPL Ga Devices marked as category 1 G or EPL Ga Devices marked as category 1 G...

Страница 13: ...screwed in it must form a flameproof joint with the housing with a minimum thread length of 8 mm A suitable mechanical stopping box must be provided in accordance with installation requirements The c...

Страница 14: ...round test voltage 500 VAC or 707 VDC between terminal and earth ground terminal 4 4 Grounding Equipotential bonding Vortex flowmeter must be connected to the equipotential bonding system of the hazar...

Страница 15: ...y only be operated if the equipment parts necessary for safety are effective in the long run and are not rendered inoperable during operation In case of operation with flammable products the measuring...

Страница 16: ...llation is within the responsibility of the operator Due to the modular design of the vortex flowmeters from a safety perspective the electrical equipment built into the signal converter can be replac...

Страница 17: ...erve the information above Also ensure that all process connections and the pipeline are depressurised and free of product Where environmentally critical products are concerned carefully decontaminate...

Страница 18: ...7 NOTES 18 OPTISWIRL 5080 www krohne com 07 2020 4008054001 AD OPTISW5080 Exd Gb Db en R01...

Страница 19: ...NOTES 7 19 OPTISWIRL 5080 www krohne com 07 2020 4008054001 AD OPTISW5080 Exd Gb Db en R01...

Страница 20: ...ces Head Office KROHNE Messtechnik GmbH Ludwig Krohne Str 5 47058 Duisburg Germany Tel 49 203 301 0 Fax 49 203 301 10389 info krohne com KROHNE 07 2020 4008054001 AD OPTISW5080 Exd Gb Db en R01 Subjec...

Отзывы: