4
MOUNTING
6
OPTISENS TSS 7000
www.krohne.com
06/2020 - 4007124802 - AD OPTISENS TSS 7000 3-A R02
4.1 Mounting
Observe the following for a measuring point in accordance with 3-A standards:
•
The wetted part materials agree with 3-A ANSI General Requirements Standard number 00-
01-2018. Please be aware that the gasket corresponds to the product and the pressure limits.
•
The device has to be installed in such a way that drainability is ensured.
•
Center the clamp or Varivent
®
connection over the corresponding fitting and gasket.
•
Use a suitable connection element (e.g. half ring or clamp ring connection) to attach the
device according to the manufacturer
´
s specification.
•
Welding connections must have a surface roughness
≤
0.8
µ
m.
•
Weld in connections must be welded flush to the inside of the tank or pipeline.
The following figure illustrates the mandatory mounting position.
If a different mounting position is chosen, the user must implement the appropriate measures to
enable the sensor to self-drain.
Figure 4-1: Mandatory mounting position
45°
0°