 MOUNTING 

OPTISENS TSS 7000

www.krohne.com

06/2020 - 4007124802 - AD OPTISENS TSS 7000 3-A R02

4.1  Mounting

Observe the following for a measuring point in accordance with 3-A standards:

The wetted part materials agree with 3-A ANSI General Requirements Standard number 00-
01-2018. Please be aware that the gasket corresponds to the product and the pressure limits.

The device has to be installed in such a way that drainability is ensured.

Center the clamp or Varivent

®

 connection over the corresponding fitting and gasket.

Use a suitable connection element (e.g. half ring or clamp ring connection) to attach the 
device according to the manufacturer

´

s specification.

Welding connections must have a surface roughness 

 0.8

µ

m.

Weld in connections must be welded flush to the inside of the tank or pipeline.

The following figure illustrates the mandatory mounting position.

If a different mounting position is chosen, the user must implement the appropriate measures to 
enable the sensor to self-drain.

Figure 4-1: Mandatory mounting position

45°

Содержание OPTISENS TSS 7000

Страница 1: ...46 04 OPTISENS TSS 7000 OPTISENS TSS 7000 OPTISENS TSS 7000 OPTISENS TSS 7000 Supplementary Instructions Supplementary Instructions Supplementary Instructions Supplementary Instructions KROHNE 06 2020...

Страница 2: ...1 1 Function 3 1 2 Scope 3 2 Sanitary Standards 4 2 1 3 A Standards Meaning and implementation 4 3 Device version 5 3 1 Device version 5 4 Mounting 6 4 1 Mounting 6 5 Maintenance and cleaning 7 5 1 M...

Страница 3: ...ortant information regarding the use of the OPTISENS TSS 7000 according to the 3 A Sanitary Standards for Refractometers and Energy Absorbing Optical Sensors for Milk and Milk Products Number 46 04 1...

Страница 4: ...ration issues a symbol licence for the 3 A logo Use of devices in accordance with 3 A The use of devices and components that bear the 3 A logo is evidence that the design has been inspected The certif...

Страница 5: ...ccording to 3 A Standard 63 04 using appropriate gaskets The device has to be installed in such a way that drainability is ensured Gaskets The user is responsible for Using a suitable material Definin...

Страница 6: ...inability is ensured Center the clamp or Varivent connection over the corresponding fitting and gasket Use a suitable connection element e g half ring or clamp ring connection to attach the device acc...

Страница 7: ...materials 3 A ANSI General Requirements Standard number 00 01 2018 Cleaning Keep the sensor clean using validated state of the art cleaning processes Check that it has been successfully cleaned by rem...

Страница 8: ...mp PP 0 85 C 32 185 F VGAY 4 72222040110x Optical path length 10 mm 3 TriClamp PP 0 85 C 32 185 F VGAY 4 72322040110x Optical path length 20 mm 3 TriClamp PP 0 85 C 32 185 F VGAY 4 72422040110x Optica...

Страница 9: ...NOTES 7 9 OPTISENS TSS 7000 www krohne com 06 2020 4007124802 AD OPTISENS TSS 7000 3 A R02...

Страница 10: ...7 NOTES 10 OPTISENS TSS 7000 www krohne com 06 2020 4007124802 AD OPTISENS TSS 7000 3 A R02...

Страница 11: ...NOTES 7 11 OPTISENS TSS 7000 www krohne com 06 2020 4007124802 AD OPTISENS TSS 7000 3 A R02...

Страница 12: ...es Head Office KROHNE Messtechnik GmbH Ludwig Krohne Str 5 47058 Duisburg Germany Tel 49 203 301 0 Fax 49 203 301 10389 info krohne com KROHNE 06 2020 4007124802 AD OPTISENS TSS 7000 3 A R02 en Subjec...

Отзывы: