6.1

 

Error messages 

 

Error message  

Cause 

Action 

Slope error 

The slope determined by calibration 
was higher than 200% or lower than 
20 %. 

Please check the sensor connection 
and cable, the flow, and the 
temperature sensor and settings.  
Then repeat the calibration. 
If the error message remains, the 
sensor has to be cleaned, regenerated, 
or replaced. 

Error input 1 

The measuring sensor gives no 
signal. 

Please check the connections, the 
cable and the sensor for signs of 
damage. 

Error input 2  

The temperature sensor gives no 
signal. 

Please check the connections, the 
cable and the sensor for signs of 
damage. This message also appears if 
automatic temperature compensation 
was selected although no temperature 
sensor was used. 

Limit 1 / 2  

The measured value exceeded  
limit 1 (or dropped below limit 2, 
respectively). 

Please check the dosing and readjust 
the control parameters, if necessary. 

Dosage check 1 / 2 

Controller 1 (or 2, resp.) gives out a 
100% output for more than the 
defined period of time. 

Please check the dosing, especially the 
feeding tubes and connections. 

Caution! Carefully check for leaking 
chemicals!

 

 

Ext. controller stop  

The digital input has been short-
circuited (in the „contact“ mode). 

This only indicates the external 
controller stop. If, however, you have 
connected a level sensor, this 
message appears due to the „lack of 
water“ alarm. 

 

No water 
 
 

The digital input has been short-
circuited (in the „flow“ mode). 

The flow has dropped below minimum. 
Check the flow at the outlet of the 
armature. If the flow is low although the 
water supply is ok, take out the filter 
and clean it with pressurized water. If 
the flow is still low, also clean the flow 
sensor. 

 

 

Содержание OPTISENS AAC 050 R

Страница 1: ...llation and Operating Instructions OPTISENS AAC 050 Measuring and control instruments for potentiostatic measurements of Chlorine total Chlorine Chlorine dioxide Oxygen Ozone and Hydrogen peroxide 07...

Страница 2: ...ion of the converter 15 4 2 Calibration 15 4 3 Temperature compensation 16 4 3 1 Calibration of the temperature measurement 16 4 4 Automatic Sensor Cleaning ASR option 17 5 Adjustments of the controll...

Страница 3: ...Manual OPTISENS AAC 050 3 OPTISENS AAC 050 9 2 1 Panel mounting enclosure 30 9 2 2 Wall mounting enclosure 30 10 Device return form 31 11 Customer settings for reference 32...

Страница 4: ...le and Delivery found on the back of the invoice and forming the basis of the purchasing contract are applicable General limitation on liability Unless otherwise expressly set forth in the Standard Te...

Страница 5: ...safe usage follow all instructions carefully and pay special attention to all warnings issued in this manual If the device is visibly damaged or has been stored inappropriately or if there are any dou...

Страница 6: ...ntroller types ON OFF controller with hysteresis P controller as Pulse Pause Impulse Frequency or steady controller PI controller as Pulse Pause Impulse Frequency or steady controller Hysteresis adjus...

Страница 7: ...r by sliding the slot under a screw which is an alternative for limited space Either way you have to fix it additionally with two screws ATTENTION Install the device in a place where it is not put und...

Страница 8: ...upon the upper screw In that case drill the upper hole 120mm 4 7 above the lower two 2 Or you can slip the fixture on the back of the device under the upper screw In that case the upper hole has to be...

Страница 9: ...ly The potentiostatic measurement is interference sensitive especially when using membrane sensors Use a special screened cable Membrane sensors are delivered complete with cable When using the relays...

Страница 10: ...lectrode sensors bridge 3 4 1 2 measurement 3 6V Membrane Sensor with electronics 1 4 4 6V Pt100 NTC 5 6 Analog output 9 10 9 10 max burden 500 Ohm Digital input 11 12 11 12 external controller stop o...

Страница 11: ...ement 3 6V Membrane sensors with electronics 1 4 4 6V Pt100 NTC 6 7 Display contrast Display Potentiometer to adjust the display contrast Analog output 11 12 11 12 max burden 500 Ohm Relay 1 14 15 Rel...

Страница 12: ...ring With key Enter changes are stored For your convenience arrows in the display indicate the directions you can take from your position in the menu 3 1 How to adjust parameters Temp comp Manual 1 Wh...

Страница 13: ...mp Temperature compensation only for pH measurements Enter password Password function Set points Controller settings set points P ranges I functions Limit values Alarm function Basic settings Basic se...

Страница 14: ...anual compensation Flow rate Display of the flow measurement Basic settings Basic settings Cal Pt NTC Calibration of the temperature measurement Flow rate Calibration of the flow measurement Flow alar...

Страница 15: ...peroxide H2O2 0 0 100 0 mg l Input 2 works with the temperature sensors Pt100 or NTC 4 2 Calibration Main menu Calibration DPD Calibration DPD Calibration 0 48 mg l Slope Enter the concentration dete...

Страница 16: ...nual Temp 25 0 C Basic settings Calibration Pt 100 NTC Basic settings Cal Pt 100 NTC Cal Pt 100 NTC 0 0 C Temp coefficient Temp coeff Temp coeff 0 0 K Select between two ways of compensation 1 Automat...

Страница 17: ...last measured value for five minutes and the message cleaning in progress is displayed As a safety measure attempts to calibrate within these five minutes are ignored Activation and timing The cleani...

Страница 18: ...eved by either reducing the switch frequency Impulse frequency controller or reducing the time within a given period of time in which the relay is ON pulse pause controller The parameters for a P cont...

Страница 19: ...nu Set points and controller version Set points Set point S1 You can choose different 1 50 mg l Controller versions P range S1 For S1 and S2 0 20 mg l Integral time S1 000 sec Set point S2 0 50 mg l C...

Страница 20: ...pulse pause controller you have to set the following parameters 1 Set points S1 and S2 S1 refers to relay 1 S2 refers to relay 2 2 P range and integral action time Adjust a P range 0 For a P controll...

Страница 21: ...y Turn on delay Turn on delay OFF ON Turn on delay 180 s 5 5 Turn on delay Set a delay time which has to pass before the controller is activated after start up or power interrupt This allows the measu...

Страница 22: ...1 402 mS 25 0 C S1 S2 Man 6 Switch ON relay 2 with key The square to the left of S2 gets dark 1 402 mS 25 0 C S1 S2 Man 7 Switch OFF relay 2 again with key The square gets light 1 402 mS 25 0 C S1 S2...

Страница 23: ...s it happens regularly that the measured value exceeds a limit for a short period of time To avoid having an alarm issued under these circumstances you can adjust a turn on delay which has to pass bef...

Страница 24: ...of calibration failure the controller is not deactivated to enable you to continue the dosing or treatment process until a replacement sensor is at hand Sensor check during measurement During measure...

Страница 25: ...e sensor was used Limit 1 2 The measured value exceeded limit 1 or dropped below limit 2 respectively Please check the dosing and readjust the control parameters if necessary Dosage check 1 2 Controll...

Страница 26: ...g 4 20 mA the resolution is lower but defective cable connections are immediately evident With the parameters Begin and End you define which part of the measuring range you want to read out Alternativ...

Страница 27: ...a Bayonet lock Turn the lock to the left until the fuse pops up Exchange it and fix the new fuse by turning the lock to the right Put the front back on and fix it tightly WARNING Disconnect the power...

Страница 28: ...tion of the device hardware and software 8 7 2 Analog inputs Here you can see the raw data the device obtains from the sensors They are not influenced by compensations or calibration and offer valuabl...

Страница 29: ...ature with dimensions indication of relay status Current outputs 2 x 0 4 20 mA galvanically isolated max loading 500 Ohm Interface option RS485 Baud rate 9600 data format 8Bit 1 Start and 1 Stop bit C...

Страница 30: ...Manual OPTISENS AAC 050 OPTISENS AAC 050 30 9 2 Dimensions 9 2 1 Panel mounting enclosure 9 2 2 Wall mounting enclosure...

Страница 31: ...firming that the device is safe to handle If the device has been operated with toxic caustic flammable or water endangering products you are kindly requested to check and ensure if necessary by rinsin...

Страница 32: ...or conductivity for Temperature Controller S1 Controller S2 Controller S1 Controller S2 Start Start End End Controller Controller S1 Controller S2 Acting direction Acting direction raise lower raise l...

Отзывы: