2.7  Ambient temperature / temperature classes

Because of the influence of the temperature of the product, no fixed temperature class is 
assigned to variable area flowmeters with attached electrical equipment. In fact, the 
temperature class of a device is a function of the temperature of both the product and the 
environment. There is no distinction between devices with one or two contacts. The assignment 
is outlined in the following tables.

The tables take into account the following parameters to determine the permissible 
temperature class:

Ambient temperature T

amb.

Product temperature T

m

attached equipment

Supply type (maximum values for P

i

)

Maximum permitted product temperature T

m

  for limit switch TG21

The maximum permitted ambient and product temperature for the type MS14 limit switch is 
85

°

C  / 185

°

F for temperature classes T6 ... T1.

The minimum permitted ambient temperature T

amb

 for all versions is -20

°

C / -4

°

F.

2.8  Electrical data

The connection of  limit switches MS14/.  or TG21 may only be connected to  intrinsically safe 
circuits
to DIN 19234 with the following max. values:

MS14/.

TG21

Supply

Type 1

Type 2

Type 3

Temperature 
class

T6

T5

T4...T1 T6

T5

T4...T1 T6

T5

T4...T1

T

m

 in 

°

C

 1

70

85

95

60

75

95

35

50

70

T

m

 in 

°

F

 1

158

185

203

140

167

203

95

122

158

1

These values may be limited by the Installation and Operating instructions. Observe the maximum values listed in the 
Installation and Operating instructions.

Pi

Ui

Ii

Ci

Li

1 W

30 VDC

100mA

 0 nF

 0 

µ

H

Supply

Pi

Ui

Ii

Ci

Li

Type 1

34 mW

16 VDC

25mA

165 nF

150 

µ

H

Type 2

64 mW

16 VDC

25mA

165 nF

150 

µ

H

Type 3

169 mW

16 VDC

52mA

165 nF

150 

µ

H

Содержание GA24

Страница 1: ...evice category II2G with electrical internals GA24 GA24 GA24 GA24 Supplementary instructions Supplementary instructions Supplementary instructions Supplementary instructions KROHNE 04 2015 4000836702...

Страница 2: ...on code 4 2 3 Marking 5 2 4 Flammable products 7 2 5 Equipment category 7 2 6 Types of protection 7 2 7 Ambient temperature temperature classes 8 2 8 Electrical data 8 3 Installation 9 3 1 Installatio...

Страница 3: ...own responsibility conformity with the protection goals of Directive 94 9 EC for use in hazardous areas with gas The EC Type Test Certificate of the Physikalisch Technische Bundesanstalt PTB forms the...

Страница 4: ...n be mounted to the on site display 2 2 Safety description code The safety description code consists of the following elements positions which are not needed are omitted no blank positions 1 R Stainle...

Страница 5: ...te GA24 1 Device type 2 Manufacturer 3 Manufacturer s website 4 Note manual 5 Ex data 6 Serial No 7 Year of manufacture 8 Year of manufacture 9 PED data 10 Design data Temperature pressure rating Name...

Страница 6: ...complies to intrinsic safety EN 60079 11 paragraph 5 7 Marking Limit switches MS14 1 Type 2 Manufacturer 3 Year made 4 Observe Operation and Installation manual 5 Ambient temperature 6 KROHNE website...

Страница 7: ...ntrinsic safety protection level ia as per EN 60079 11 The marking is II 2G Ex ia IIC T6 T1 Gb II 2G Ex ia IIC T6 T1 Gb II 2G Ex ia IIC T6 T1 Gb II 2G Ex ia IIC T6 T1 Gb CAUTION Operation with flammab...

Страница 8: ...permitted product temperature Tm for limit switch TG21 The maximum permitted ambient and product temperature for the type MS14 limit switch is 85 C 185 F for temperature classes T6 T1 The minimum per...

Страница 9: ...erved Variable area flowmeters must be installed in such a way that There is no danger from mechanical impact effects There are no external forces affecting the indicator part The device is accessible...

Страница 10: ...ust be selected according to prevailing installation standards e g EN 60079 14 The outer diameter of the connecting cable must be within the sealing range of the cable entry The connecting cables must...

Страница 11: ...immediately afterwards 5 3 Electrostatic charge In variable area flowmeters it is possible under field conditions for charge separation to occur in the measuring tube due to the transport of non condu...

Страница 12: ...lean the meter in accordance with the installation and operating instructions for non explosion proof versions of the product The measuring unit must be removed for cleaning To do this follow the inst...

Страница 13: ...NOTES 7 13 GA24 www krohne com 04 2015 4000836702 MA G24 Ex II2G AD R03 en...

Страница 14: ...7 NOTES 14 GA24 www krohne com 04 2015 4000836702 MA G24 Ex II2G AD R03 en...

Страница 15: ...NOTES 7 15 GA24 www krohne com 04 2015 4000836702 MA G24 Ex II2G AD R03 en...

Страница 16: ...nalysis products Products and systems for the oil gas industry Measuring systems for the marine industry Head Office KROHNE Messtechnik GmbH Ludwig Krohne Str 5 47058 Duisburg Germany Tel 49 203 301 0...

Отзывы: