SECTION 3
IEEE-488 STD (GPIB) PROGRAMMING
3. 1
INTRODUCTION
The Model 3945 remote programming interface accepts both ASCII data command and IEEE-488
standard commands (ATN true) for control of the unit.
In presenting the information required to program the Model 3945 via the IEEE-488 STD bus, this
manual presupposes a user knowledge of both ASCII data and IEEE-488 bus commands.
3. 2
PRELIMINARY PROGRAMMING INFORMATION
3.2.1
GPIB Primary Bus Address
The GPIB primary address and software line-termination-character-sequence (LTCS) selection is set
via the front panel keyboard as listed in Tables 3.1 and 3.2. These two parameters are stored in
non-volatile memory and will be remembered indefinitely, even when the power to the unit is re-
moved. They do not need to be reentered each time the unit is turned on.
The LTCS affects the GPIB in the TALKER mode only (data output from the 3945 to the GPIB). After the
printable characters have been sent, non-printable characters, such as carriage return (CR) and line feed
(LF), are often required to achieve the desired results in various computers. Table 3.2 lists the various key
sequences with the LTCS it selects.
SETTING AND DISPLAYING THE GPIB PRIMARY ADDRESS
Function
Keyboard Entry
a.
To set a primary address from 0 to 30
[x][SHIFT][MEGA]
b.
To display the primary address
[SHIFT][MEGA]
LINE-TERMINATION-CHARACTER-SEQUENCE
Line-Termination Character-Sequence
Keyboard Entry
a.
None (EOI only)
[0][SHIFT] [ALL CH]
b.
Carriage return (with EOI)
[1][SHIFT] [ALL CH]
c.
Line Feed (with EOI)
[2][SHIFT] [ALLCH]
d.
Carriage return followed by line feed (with EOI)
[3][SHIFT] [ALL CH]
e.
Line feed followed by carriage return (with EOI)
[4][SHIFT] [ALL CH]
f.
Display present LTCS
[SHIFT] [ALL CH]