37
The gage should be verified before each series of measurements by using a setting ring.
If there is a deviation from the specified measurement the device should be re-adjusted
via SETUP-Menu
OFFSET
.
Always the total selected OFFSET value will be shown!
The value by which the gage must be adjusted has to be added or
subtracted.
In order to increase the accuracy of the measurement, the gage should
be calibrated in the same position in which later the measurement will
be effected.
The instrument must be switched on.
The SETUP – Menu can be activated in each
program.
1
1x long
Activating the SETUP – Menu
2
1
x short
„
OFFSET
“ will appear in the display
3
1x long
Activating the OFFSET – Menu
4
short
Selecting the sign
plus (+) / minus (-)
Press key several times
5
1x
short
Selecting the first digit to be settled
6
short
Increasing the first digit or…..
7
short
…..Reducing the first digit
8
1x
short
Selecting the next digit
9
Repeat step 6-8 until the OFFSET value is totally entered.
For scale interval 0,005 3 digits are editable!
10
1x long
For confirming and saving
11
Afterwards you will return to the SETUP -
Menu.
Nullpunkt-Korrektur (OFFSET)
Zero point adjustment (OFFSET)