11
The gage should be verified before each series of measurements by
using a gage block.
If there is a deviation from the specified measurement the device
should be re-adjusted by using the rotating scale (E).
The measurements have to be repeated until the gage is
correctly adjusted.
In order to increase the accuracy of the measurement, the gage should be
adjusted in the same position in which later the measurement will be
effected.
The gage provides tolerance marks (D) that can be set to an acceptable tolerance range
which allows a quick measurement reading.
Part – No.:
1070-11
Covering glass (E)
1070-12
1 pc. Tolerance marks (D)
1070-03
Cover (G)
mm – gages
inch – gages
A1
*)
B1
*)
Holding unit
Wooden box
D110
D610
0251-04
0251-04
8004-50
1732-45
D110T
D610T
1533-51
0701-77
D110S
D610S
–
–
D1R10
D6R10
0251-04
–
D1R10S
D6R10S
–
–
D220
D720
0251-04
0251-04
D220T
D720T
1533-50
0701-75
D220S
D720S
–
–
D2R20
D7R20
0251-04
–
D2R20S
D7R20S
–
–
*)
A1
Measuring contact (movable measuring arm)
B1
Measuring contact (fixed measuring arm)
–
Measuring contacts are not interchangeable
Tolerance marks
Spare parts / Accessories
Display adjustment
E
D