7 Preservation
6.5 Storing the flowmeter
Fig. 8: Fastening of hoisting equipment - principle diagram
1.
Fasten hoisting equipment on the flowmeter.
2.
Transport the flowmeter to the installation location or temporary storage site.
6.5 Storing the flowmeter
As a result of the calibration, the internal components of the flowmeter are wetted with calibration liquid
that has a preservative effect. In addition, a special anticorrosive agent is sprayed onto the interior of
the flowmeter before being dispatched. The connections of the flowmeter are fitted with protective cov-
ers. The preservation applied at the factory will protect the flowmeter for up to six weeks, if it is stored
in a dry and clean location. The manufacturer optionally offers a long-term preservation for storage
times of up to 60 months. The flowmeter is additionally packed in hermetically sealing anti-corrosion
paper.
ATTENTION
Damage to equipment and corrosion if stored improperly and during longer standstills.
► Protect the flowmeter against damage, heat, sunlight, dust and moisture.
► Protect against corrosion during longer standstill.
► Observe measures for storing and preservation.
1.
Store cool and dry and protect against sunlight.
2.
Observe the information on preservation.
7 Preservation
7.1 Preservation table
Preservation has to be carried out additionally under the following conditions:
Type of delivery
Condition
Standard delivery
o
Storage time exceeding six weeks
o
Unfavourable storage conditions such as high humidity,
salty air, etc.
Delivery with long-term preservation
o
Opened or damaged packaging
Tab. 7: Conditions for additional preservation
Notice
After a longer storage time the manufacturer recommends that you have the flowmeter recalibrated
Ä
Operating instructions
OIO 08en-GB Edition 2020-01
21