RB-525-B04

30

6.10 Power Sa ving Function

Power saving function will start operating when switch operation is suspended with the power on.

(The switchover time (in minute) can be selected at '

SAVE

' on the mode selection screen. )

To return to the measurement mode, press any switch (any switch on the front panel or the measurement

start switch).

Measurement Mode

Power Indicator (PW)

 ON

Power Saving Condition

Power Indicator (PW)

flashing

Return to Measurement Mode

Power Indicator (PW) 

ON

Leave idling for a setting time

Press any switch

6.11 Output Te rminal

Video Terminal

This terminal outputs an NTSC video signal.

If you connect an external monitor such as a portable TV or similar unit to KW-2000 with pin plug (video)

cable, you can observe and check the same image that appears on the internal monitor screen of KW-2000

simultaneously.

Содержание KW-2000

Страница 1: ...RB 525 B04 Operations Manual Auto Refkeratometer...

Страница 2: ...t SAFETY CONSIDERATION KW 2000 is a Class I Type B medical instrument as well as LED Class 2 product This instrument complies with Medical Device Directive 93 42 EEC A great deal of consideration has...

Страница 3: ...trical ground safety ground at the power outlet Personal injury may result form electric shock etc If the instrument fails to work properly you should not try to repair the fault Consult your dealer i...

Страница 4: ...6 5 Kerato Peripheral Measurement 16 6 6 IOL Measurement 18 6 7 Menu Screen Setting 19 6 7 1 Each Item Description 20 6 7 2 Optional Functions 23 6 8 Auto Start Function 27 6 9 Data Screen Function 2...

Страница 5: ...000 sheets Chine rest liner pin 2 Dustproof cover 1 Operations manual 1 Use accessories specified by us to avoid any malfunction or failure Extra care should be taken for storage of a model eye Avoid...

Страница 6: ...al interface connector External monitor output terminal Power plug connector Fuse holder Power switch Joystick Measurement start switch Rating plate Printer LCD monitor Body Base Anti sliding lock IOL...

Страница 7: ...ents below for installation Temperature 10 C 40 C Relative Humidity 30 85 3 Keep away from inflammable or explosive gases as well as storage areas housing medical supplies and chemicals 4 Avoid instal...

Страница 8: ...tral cleanser For disinfecting them especially where the examinee may contact hydrogen peroxide Oxydol is recommended 6 If the instrument is not used for any length of time remove the power cord from...

Страница 9: ...with the chin rest liner pins on the chin rest Refer to the figure on the right Always use the chin rest liners following above For sanitary consideration disinfecting the chin rest with Oxydol is rec...

Страница 10: ...the power is OFF 2 Insert the power cord into the instrument s power plug connector Then insert the plug into a general purpose outlet 3 Turn the power switch ON I O POWER T 2A 250V I O POWER T 2A 25...

Страница 11: ...ng 0 10 12 13 5 and 15mm Indicates a mark for a cylindrical value Indicates an eye currently measured Refractive measurement result is displayed in this area Indicated values are S C and A Keratometry...

Страница 12: ...Measurement mode selection switch Print switch IOL switch Power indicator It blinks when the saving function is activated R K MENU Down Backward Right Up Forward Left Operation of Joystick Forward th...

Страница 13: ...on LCD monitor carry out alignment for correct measurement 5 Press the measurement start switch when the alignment mark reaches a center of the reticle mark and take measurement Refer to 6 2 Alignment...

Страница 14: ...into the center of the reticle mark 3 Keeping the alignment mark in the center of the reticle mark focus on the subject eye and press the measurement start switch For auto measurement measurement will...

Страница 15: ...easurement result Axis Angle Near pupillary distance Radius of Curvature minmum Refractive measurement result Spherical value Cylindrical value Axis angle Radius of Curvature maximum PC result is indi...

Страница 16: ...etry data Refractive data Right eye data Sample printout when PRINT FORM is set for ALL Message area Optimum value indicated when each eye is measured more than three times Left eye data Pupillary dis...

Страница 17: ...sition becomes blue Operation Measurement guiding mark Measurement position Normally measurement is taken in the following order C S I N C Central center S Superior top T Temporal temple side I Inferi...

Страница 18: ...orneal Center only When R K or Kerato measurement is also taken only optimum values are printed R mm D AX R1 7 59 44 50 120 R2 7 57 44 50 107 AVE 7 58 44 50 CYL 0 00 S 7 84 V 7 51 T 7 66 H 7 55 N 7 71...

Страница 19: ...12 Ker Ref 0 25 0 0 R R K MENU S C A In IOL measurement mode the IOL switch icon S C and A change to purple IOL mode measurement screen Printout sample Measurement results of IOL measurement mode are...

Страница 20: ...LIGHT BRIGHT NORMAL DARK SCREEN ADJ SAVE min OFF 3 5 10 BUZZER HIGH LOW OFF DATA CLEAR s OFF 1 2 3 OPTION MESSAGE No RS232C DATE FORM YMD DMY MDY DATE 2002 11 20 TIME 14 29 22 the First Screen the Sec...

Страница 21: ...surement start switch is pressed once QUICK A set number of measurements are continuously taken when the measurement start switch is pressed once A maximum of ten measurements is available A target mo...

Страница 22: ...en SAVE min Selects switchover time in minute to activate power saving function BUZZER Sets volume of buzzer at measurement HIGH turns volume up LOW turns volume down OFF no buzzer DATA CLEAR S Sets t...

Страница 23: ...e YMD year month date DMY date year month MDY month date year DATE Sets and corrects date TIME Sets and corrects time If you wish to correct date or time move the cursor to the number you wish to chan...

Страница 24: ...e the input cursor to END and press switch to go back to the menu screen When you need to change the character already inputted press switch to move the input cursor to the character you wish to input...

Страница 25: ...er OFF no print out ON prints out the number Selects monitor display setting of examinee s number OFF no display ON displays on the screen No SET 00001 RESET PRINT OFF ON DISPLAY OFF ON 1 Move the cur...

Страница 26: ...38400bps 19200bps 9600bps 4800bps 2400bps 2 CHARACTER selection of data bit CHARACTER option Standard setting 8 7 3 PARITY selection of transmission data check PARITY option Standard setting EVEN ODD...

Страница 27: ...DTR 20 6 DSR If you have any inquiries about changes in transmission parameters and or detail connections please direct them to the relevant agent or distributor BAUDRATE 9600 CHARACTER 8 PARITY EVEN...

Страница 28: ...ft or right side you have measured will disappear In addition an arrow to urge the shift to the other eye will appear and flash in the center of the screen 5 Measure the other eye in the same manner T...

Страница 29: ...e to measure with auto start function realignment is necessary When you wish to cancel auto start function change the setting of START back to MANUAL in the menu screen NOTE The data printed out depen...

Страница 30: ...75 0 25 62 4 75 0 00 4 75 0 00 4 75 0 00 4 75 0 00 4 75 0 25 62 4 75 0 00 4 75 0 00 4 75 0 00 4 75 0 00 4 75 0 00 mm D AX R1 7 59 44 50 120 R2 7 57 44 50 107 AVE 7 58 44 50 CYL 0 00 S 7 58 T 7 66 V 7...

Страница 31: ...asurement start switch Measurement Mode Power Indicator PW ON Power Saving Condition Power Indicator PW flashing Return to Measurement Mode Power Indicator PW ON Leave idling for a setting time Press...

Страница 32: ...position for the examinee 5 When the eyelash or eyelid interfere measurement error will occur in measurement Urge the examinee to keep his her eye wide open 6 Tear residue or eye mucus etc trapped on...

Страница 33: ...spherical measurement range 25D to 25D when VD 0 CYL OVER Exceeded cylindrical measurement range 0 to 10 Motor fault Detected abnormality in motor control system EEPROM fault Failed to initialize Cut...

Страница 34: ...a contact lens holder of the model eye as following 1 Put a small amount of water on a concave side of the contact lens holder 2 Place the contact lens so that its convex side faces the holder Contac...

Страница 35: ...ctivate the saving function The main body can not be moved in a horizontal direction Anti sliding lock may be locked Unscrew the lock under the base The moving parts such as a joystick are not moving...

Страница 36: ...r through the printer cutter slot and lower the head up lever and the printer cutter back to the original positions 6 Draw the paper a cutter slot of a printer cover and reattach the printer cover to...

Страница 37: ...e power cord from the outlet Lower the optical unit to the bottom original position Secure the body with the anti sliding screw lock Put a dustproof cover on the optical unit 2 Notes on storage enviro...

Страница 38: ...cal Inspection and Maintenance To prevent malfunction and accidents and maintain the performance and reliability of the product it is recommended to request your distributor for the periodical inspect...

Страница 39: ...ratometry measurement approx 0 07 sec approx 0 07 sec Printer Thermal line printer with automatic cutter paper width 57mm Internal Monitor 5 6 inch LCD monitor color Shifting Range for Sliding Body ba...

Страница 40: ...3279 7312 Facsimile 81 3 3245 1109 20001 So Vermont Ave Torrance CA 90502 U S A Phone 1 310 327 1913 Facsimile 1 310 327 4177 Immermannstrasse 65A 40210 Dusseldorf F R Germany Phone 19 211 35 3444 45...

Отзывы: